Model Number 8032
Medical Device Problem Code Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/11/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A follow-up report will be submitted upon completion of our investigation.
 
Event or Problem Description
It was reported by the customer that they had identified an issue with a recent batch of chest tubes received.The ends of the chest tubes were oval-shaped on the distal end rather than the standard round shape.The oval shape was not consistent with the required specifications needed for the patients and the facility was concerned about kinking/blockages that could occur in the patient and affect clinical performance.Note: it was reported this chest tube was noticed from the hospital's inventory, prior to use and there was no patient involvement.
 
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
Additional fields: h3, h6 related mdr¿s: 3011175548-2025-0000093, 3011175548-2025-0000097, 3011175548-2025-0000096 & 3011175548-2025-0000095 the customer reported that the end of the 32 fr straight pvc catheters was found to be oval instead of round which caused a kink and blockage in one catheter what was found during use.The user reported that they observed four (4) additional catheters within their inventory with the same oval shape at the end of the catheter.The additional four (4) catheters were not used.The user provided a picture which does show the described issue.It appears as if the catheter was compressed in some way and does not appear otherwise damaged.The user originally had planned to return the unused catheters but changed their mind because another clinician had been able to use the catheters without issue.The lots of these catheters were manufactured between january 22, 2025 and may 13, 2025.Due to the wide manufacturing timeframe of these lots and the lack of any other complaints of this issue, as well as the lack of any inventory from in or around that time-frame contemporaneous samples will not be evaluated.Information about the product storage was not received.No other complaints of this issue have been received and there have been no ncrs for this issue from incoming inspection or manufacturing.Therefore, the most likely cause of the problem lies within the user's facility.The manner in which the catheters are stored may compress them, leading to the shape deformation.A dhr review was completed and did not identify any anomalies in manufacturing.No evidence has been identified to indicate that manufacturing or design contributed to this complaint.A recurring lot number query was conducted for each reported lot number and there have been no other complaints associated with the production lot of finished goods except for the ones stated in the reported complaints.The ifu provides adequate instructions for the setup and use of the device.Complaint trending found that the actual occurrence level did not exceed the anticipated occurrence level.Two excursions were identified, triggered by these complaints, for which (b)(4) was initiated.A complaint history review was completed which found no similar complaints.This complaint is confirmed by the provided picture.It cannot be confirmed that the device was nonconforming when it left the manufacturing facility.The evidence indicates that the most likely cause of the complaint is the storage conditions/technique at the user facility.The root cause of these complaints is impossible to define.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PVC CATHETERS STRAIGHT
Common Device Name
CATHETER, CHOLANGIOGRAPHY
Manufacturer (Section D)
ATRIUM MEDICAL CORPORATION
40 continental blvd
merrimack NH
Manufacturer (Section G)
ATRIUM MEDICAL CORPORATION
40 continental blvd
merrimack NH
Manufacturer Contact
lori gosselin
40 continental blvd
merrimack, NH 
MDR Report Key22771146
Report Number3011175548-2025-000009
Device Sequence Number9621747
Product Code GBZ
UDI-Device Identifier20650862130328
UDI-Public(01)20650862130328(17)271127(10)ME242066
Combination Product (Y/N)N
Initial Reporter StateDE
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 09/12/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number8032
Device Catalogue Number8032
Device Lot NumberME242066
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/18/2025
Supplement Date Received by Manufacturer09/08/2025
Initial Report FDA Received Date08/11/2025
Supplement Report FDA Received Date09/14/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/22/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
N/A.
Patient SexUnknown
-
-