| Model Number |
UNI GLENOID-PERIPHERAL LOCK SCRW |
| Medical Device Problem Code |
Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
07/23/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Event or Problem Description
|
|
It was reported that during an anatomical shoulder prosthesis with universal glenoid, the base plate could not be implanted completely.The surgeon replaced the peripher and central screw without success.Per complaint reporter, there was no harm for patient, operator or third party.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.***update dw 31-jul-2025: further information was provided that the reported base plate was initially implanted four years ago.It was further reported that the patient underwent revision surgery 3 months ago where the inlay ar-9121-02 was revised.The patient continued to have issues and underwent the second revision on (b)(6) 2025.During the revision, the initially reported event has been identified and the surgeon revised the base plate and 3 screws.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6.Since the device was not received, an evaluation on the product could not be performed and the reported event cannot be verified.The most likely root cause may be attributed to a patient-specific event, potentially involving bone remodeling or degradation over time, device wear, or loosening from the initial implantation, as well as anatomical changes following the initial surgery.
|
| |
|
Search Alerts/Recalls
|