|
Manufacturer ref# (b)(4).Information regarding patient weight, height, medical history, race, and ethnicity was not reported.Section d2b ¿ procode: krd/hcg.Section d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.Section h4: the device manufacture date is not known as the device lot number is not available / not reported.Since no device has been received for analysis, no product investigation can be performed and the reported failure could not be evaluated.The lot number is not known; therefore, a device history record review cannot be completed.With the information available and without the product available for analysis, the reported customer complaint could not be confirmed.The exact cause of the event could not be determined; however, there are circumstances of the procedure that may have contributed to the reported failure.The instructions for use (ifu) contain the following recommendations: 1.Remove the detacher from the protective packaging and place it within the sterile field.The detacher is packaged separately for single patient use only.2.Confirm again under fluoroscopy that the coil alignment marker of the delivery tube creates a ¿t¿ with the proximal marker of the microcatheter.3.Verify that the rhv is firmly locked around the delivery tube to ensure that the coil does not move during connection process.Ensure the delivery tube is straight between the rhv and the detacher.4.Hold the delivery tube approximately 3 inches from proximal end and bring the detacher over the delivery tube until a firm stop is felt.5.Maintain the delivery tube position within the detacher and use a thumb to pull back the slider on the detacher.6.Continue to pull on the slider until end of travel or a click is heard.Gently release the slider until it returns to starting position or a second click is heard.Remove detacher by slowly withdrawing it from the delivery tube.Note: detacher must remain in sterile field until the procedure is completed.7.Fluoroscopically verify that the coil is detached by gently pulling back the delivery tube approximately 1cm.If the coil has detached, remove the delivery tube from the microcatheter.8.If the coil has not detached, repeat steps 3-7.If 2 unsuccessful detachments occur, discard the detacher and proceed to the manual break section.9.Repeat steps 1-8 until the procedure is completed.Discard the detacher.Caution: always perform fluoroscopic verification of detachment prior to withdrawing the delivery system fully.Failure to do so may lead to an embolic complication.10.After detaching the coil, remove the delivery tube from the microcatheter and discard.11.Repeat the above sequence for all additional coils until the procedure is complete.As part of the post market surveillance program, information from this complaint is trended to identify statistical signals for consideration of further correction action.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.Missing information from this report is identified as blank; this information was not provided in the reported event or available at the time of report submission.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by cerenovus, or its employees that the report constitutes an admission that the product, cerenovus, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
A healthcare professional reported that during a tumor embolization, a freeform mini 1mm x 2cm coil (mcr091020, (b)(6)) did not deploy.Loose metal fragment at distal end of pusher.The surgery was delayed due to the reported event.The procedure was successfully completed.There was no patient injury reported.Additional event information received on 10-aug-2025 indicated that after speaking with the account, this was probably more user error.The coil did not detach when they tried to use the mechanical handle so they attempted to manually break it.But when they did that, they did not break it in the correct manual breakpoint.They attempted to manually break it more towards the end of the proximal wire.The manual breakpoint was inside the micro catheter since they were using a headway duo 160cm micro catheter.No additional information is available.
|