| Model Number |
FIBERTAG TIGHTROPE II WITH INTERNALBRACE |
| Medical Device Problem Codes |
Break (1069); Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/23/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Event or Problem Description
|
|
On 07/23/2025, a sales representative reported via email that an ar-1588rtt2-ib fibertag tightrope ii implant had the tightrope break during basic passage.A replacement ar-1588rtt2-ib fibertag tightrope ii implant device was used to complete the case.This was discovered during a knee scope with acl procedure on (b)(6) 2025, with no reported patient harm.Additional information has been requested on 07/30/2025.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: b5, g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error of the device due to applying excessive mechanical forces upon insertion.Dfu-0147-eo: excessive force on the shortening suture strands and/or tensioning the shortening suture strands not in-line with the bone socket may break the strands and impair ability to fully seat the implant.No additional force on the shortening suture strands is required when the graft/wedge construct reaches the desired position in the femoral socket, and the graft stability is verified by pulling distally on the graft.The complaint allegation could not be confirmed as the device was not received for investigation.
|
| |
|
Event or Problem Description
|
|
Additional information was received on 07/31/2025: the broken implant was retrieved from the patient.There was a 15-minute delay, with an additional 15 minutes of anesthesia administered to the patient.No adverse effects were reported on or after surgery.
|
| |
|