MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ SINGLEPIECE IOL WITH ULTRASERT DELIVERY SYSTEM; LENS, GUIDE, INTRAOCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ SINGLEPIECE IOL WITH ULTRASERT DELIVERY SYSTEM; LENS, GUIDE, INTRAOCULAR Back to Search Results
Model Number AU00T0
Medical Device Problem Codes Failure to Advance (2524); Device Damaged by Another Device (2915); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/16/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A product was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is:(b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
 
Event or Problem Description
A health care professional reported that during an intraocular lens (iol) implant procedure, plunger bypassed the iol also when advancing the iol the anterior haptic was stretched.The bulb was also tilted to the right and the lens was partially squeezed.There was no patient contact and the surgery was completed on the same day using a backup iol.Additional information was received, all the events happened in the same surgery.
 
Additional Manufacturer Narrative
The device with the lens was returned.The plunger lock and lens stop have been removed.The plunger was oriented correctly.Viscoelastic was dried in the device.The plunger has advanced the lens to mid-nozzle.The plunger was correctly contacting the trailing optic edge.The leading haptic was straight.The trailing haptic was folded onto the anterior optic surface.It is unknown if the plunger had been retracted and advanced a second time.The nozzle was removed and cleaned for further evaluation.The lens was removed during cleaning.Topcoat dye stain testing was conducted with acceptable results.A device history record review and a non-conformance review of the reported lot number was conducted.No deviations were identified and all corresponding production release specifications defined in the device master record were met.A non-qualified viscoelastic was used.The reported lens squeezed was observed as lens stuck in device.The root cause may be related to a failure to follow the instructions for use (ifu).A non-qualified viscoelastic was indicated.The ifu instructs: during device preparation and implantation of the company intra ocular lens (iol) with the company preloaded delivery system, an company qualified ophthalmic viscoelastic device (ovd) should be used.The use of an unqualified ovd may cause damage to the lens and potential complications during the device preparation and implantation steps.Lens may become stuck in the device: due to the use of a non-qualified viscoelastic.Material properties of non-qualified ovds may contribute to underfill, overfill, misfolding of the haptics, or other inconsistent folding outcomes.If the operating room temperature is too high (greater than 23 degree centigrade / 73 degree fahrenheit) lens folding consistency is negatively affected, the lens is more adherent and this may inhibit lens advancement.Any of the above listed causes alone, or in combination, may create the reported event.Top coat dye stain testing was conducted with acceptable results.The reported plunger override was not observed was not observed upon return.It is unknown if the plunger had been retracted and advanced a second time.The reported straight leading haptic was observed.The plunger, lens and haptic positions were acceptable.The leading haptic was straight.Straight leading haptics are not a product malfunction.This is an acceptable position per the diagrams provided in the ifu.The customer indicated the use of a non-qualified viscoelastic which could be a contributing factor.A straight leading haptic may occur: due to the normal folding variations as indicated in the ifu diagrams.If the initial plunger advancement is faster than the recommended target, the leading haptic may be forced past the internal folding feature.If there is excessive delay between the device preparation and subsequent delivery the leading haptic may start to unfold.Due to the use of a non-qualified viscoelastic.Material properties of non-qualified ovds may contribute to underfill, overfill, misfolding of the haptics, or other inconsistent folding outcomes.If the operating room temperature is too high (greater than 23 degree centigrade / 73 degree fahrenheit) lens folding consistency is negatively affected, the lens is more adherent and this may inhibit lens advancement or contribute to incorrect haptic folding.Any of the above listed causes alone, or in combination, may create the reported event.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ SINGLEPIECE IOL WITH ULTRASERT DELIVERY SYSTEM
Common Device Name
LENS, GUIDE, INTRAOCULAR
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key22772391
Report Number1119421-2025-02224
Device Sequence Number13133426
Product Code KYB
UDI-Device Identifier00380652394802
UDI-Public(01)00380652394802(17)280211(11)250212(10)15989813
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/05/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberAU00T0
Device Lot Number15989813
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/22/2025
Supplement Date Received by Manufacturer08/15/2025
Initial Report FDA Received Date08/11/2025
Supplement Report FDA Received Date09/05/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/12/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BAUSCH AND LOMB EYEFILL; UNSPECIFIED REPLACEMENT LENS
Patient SexUnknown
-
-