• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON, S.A. ECLIPSE SMARTSLIP; NEEDLES, HYPODERMIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON, S.A. ECLIPSE SMARTSLIP; NEEDLES, HYPODERMIC Back to Search Results
Catalog Number 305760
Medical Device Problem Code Connection Problem (2900)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/24/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H.3.A follow up mdr will be submitted if additional information, a device evaluation, or a device history review is completed.
 
Event or Problem Description
It was reported that bd needle eclipse needle/syringe connection problem.The following information was provided by the initial reporter: faulty design of the 5/8 and the vaccine leaked due to poor connection to the syringe.The design of these needles is very poor quality.They require a great deal of force to get the needle to secure onto a syringe with a luer-lok.Level of harm: no effect - did not reach patient/person -- detected before use or does not touch patient during use.
 
Additional Manufacturer Narrative
A device history record review was completed for provided material number 305760 and lot number 2011006.The review did not reveal any abnormalities detected during the production process that could have contributed to the reported defect, and all quality tests were found to be within specification.To aid in the investigation of this issue, one (1) physical sample was received for evaluation by our quality team.Through examination of the sample, it was observed that the snap clip was activated.The needle was clogged with the medication used, as the light could not be seen through the needle itself, so movement through the needle could not be performed.Based on the returned sample, we could observe that the snap clip had been connected.The snap clip is the part of the needle that assures a proper connection with the syringe has been made.Based on the sample results, no manufacturing related defects were identified, and an exact cause could not be determined for the reported issue.It is possible that the handling of the product had an implication in the reported event.Based on the preventive measures in place, we believe the probability of this defect is very low with an unlikely chance of recurrence.At this time, further action has not been determined necessary.Our quality team will continue to closely monitor the manufacturing process for signs of this potential defect and any emerging trends.
 
Event or Problem Description
No additional information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ECLIPSE SMARTSLIP
Common Device Name
NEEDLES, HYPODERMIC
Manufacturer (Section D)
BECTON DICKINSON, S.A.
cr mequinenza
s/n
fraga, huesca 22520
SP  22520
Manufacturer (Section G)
BECTON DICKINSON, S.A.
cr mequinenza
s/n
fraga, huesca 22520
SP   22520
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key22777953
Report Number3002682307-2025-00098
Device Sequence Number9167035
Product Code FMI
UDI-Device Identifier00382903057603
UDI-Public(01)00382903057603(11)201104(17)251031(10)2011006
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K100209
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Risk Manager
Type of Report Initial,Followup
Report Date (Section B) 11/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2025
Device Catalogue Number305760
Device Lot Number2011006
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/25/2025
Supplement Date Received by Manufacturer11/17/2025
Initial Report FDA Received Date08/11/2025
Supplement Report FDA Received Date11/20/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/04/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-