It was reported that a 51 year-old caucasian female patient underwent a breast augmentation revision with a 345cc mentor memoryshape breast implant and experienced implant site calcifications on the right side post-operatively, which was diagnosed with a mammogram.As a result, the patient underwent explantation on (b)(6) 2025, and replaced with a 440cc mentor memorygel boost breast implant (catalog number shpb440).
|
|
On july 15, 2025, the mentor failure analysis lab has received the device for evaluation.On july 31, 2025, the evaluation for the device was completed.Device evaluation summary: during visual evaluation no apparent damage or visual anomalies were observed on the returned device.Calcification and breast imaging calcification of a biomaterial occurs when calcium salts are deposited in the tissue capsule surrounding the implanted device.Calcification occurs in association with a wide variety of implanted prostheses, including breast implants, heart valves, vascular grafts, and soft contact lenses, as well as in the mature breast tissue of women that have not undergone breast surgery.Calcification of the fibrous capsule that may occur following implantation of breast prostheses is generally not of clinical significance, although it may exacerbate symptoms of capsular contracture.Concern has been raised, however; that calcification may interfere with tumor detection, because microcalcifications are considered a halfmark of malignant breast disease.Although clinicians have recommended pre- and post-surgical mammograms for augmentation patients to assist in distinguishing post-operative findings from calcification associated with malignancy, benign calcification resulting from surgery is generally considered distinguishable from malignant-type calcification.Further, no published reports were identified that document any actual occurrences of missed or delayed diagnoses attributable to capsular calcification.Calcification is not a phenomenon unique to breast implants.It occurs in 11 to 53 percent of women who underwent breast reduction procedures (abboud et aj.1995, m&nick et al.1990, brown et al.1987).In the breasts of women who have not undergone any breast surgery, the prevalence of benign calcifications increases progressively with age from 8 percent in women 25 to 29 years old up to 86 percent in women 75 to 79 years old (stomper et al.1996).As part of mentor's quality process, all devices are manufactured, inspected, and released according to approved specifications.Based on the results of this investigation, it was determined that the product met the manufacturing release criteria: since no malfunction was observed during the investigation.No corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: calcification.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|