The product was returned to mentor for evaluation.Mentor conducted a visual inspection of the device.During visual evaluation no apparent damage or visual anomalies were observed on the returned device.Regarding the reported capsular contracture condition, there are not enough details to determine the factors that may have caused this condition.Capsular contracture in the patient¿s breast might be the result of the body¿s individual physiological response to the implantation of a foreign object in soft tissue.Further, capsular contracture is a known complication associated with these devices and is referenced in our current product insert data sheet.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: capsular contracture.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.Manufacturer¿s reference number: (b)(4).
|
|
It was reported that a patient underwent a breast augmentation revision surgery with implantation of a 560cc mentor smooth round high profile saline breast implant prosthesis.Post-operatively, the patient was diagnosed with left breast baker grade iv capsular contracture by a medical professional using the patient's symptoms.As a result, the patient underwent left-sided exchange with a mentor saline implant on (b)(6) 2025.
|