The reference: (b)(4) has been allocated to this case by rayner.The event description provided states that post-operatively the patient refractive outcome is not as expected.Plano was targeted and the patient's refraction is currently -2.50 / -0.50 x 087.The patient underwent an additional surgery whereby the rayone aspheric rao600c lens was explanted and exchanged.The healthcare facility states that the patient has made a complete recovery."deviation from target refraction" and "iol replacement or extraction" are listed in the "adverse events" section of the rayone ifu.There are many factors which may influence the post-operative outcome of the patient including but not limited to; incomplete removal of ovd following iol implantation (capsular block/capsular bag distension syndrome), dry eye during biometry measurement, incorrect product selection, aniseikonia combined with unsuitable lens power selection, wrong lens power caused by incorrect biometry readings/calculations (e.G., previous lasik procedure), incorrect power selection (not using optimised a-constants may be a contributory factor in this scenario - surgeons must always expect to personalise their own constants based on initial patient outcomes, with further personalisation as the number of eyes increases.), posterior synechiae, irregular capsular bag contraction, patient medical history (e.G., glaucoma, pseudoexfoliation syndrome), lens decentration or dislocation, incorrect or unstable fixation within the capsular bag, post-operative rotation of the lens, lens does not fit anatomy of the eye (e.G., too large or too small), capsulorhexis diameter too large and induced surgical astigmatism.The device has not been returned to rayner.Rayner iols undergo 100% optical inspection and power testing during the manufacturing process.All lenses tested within tolerance of the stated power (+25.5 d).Our review of production records for the rayone aspheric rao600c batch: 104253693 showed that all manufacturing and quality checks were conducted with successful results.All devices released for distribution from this batch were within tolerance, met specification criteria and were without defects.There is insufficient evidence and information available to determine the root cause of the deviation in the patient's post-operative refraction.
|