| Model Number |
406849 |
| Medical Device Problem Code |
Flaked (1246)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/13/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
This report is to advise of skiving found in the introducer during analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
One 8.5f swartz braided introducer sheath and dilator were received for evaluation.Resistance was noted near the distal end of the dilator when a brk needle/stylet assembly from current inventory was advanced through the returned dilator and sheath.The dilator tubing was cut lengthwise and a build-up of skived material was noted at the distal end of the dilator.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Brk needle instructions for use (ifu) states, ¿do not alter this device in any way.¿ the cause of the skiving is consistent with not following the instructions for use.
|
| |
|