• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATUN PRESSFIT STR STEM12X110MM; KNEE STEMS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATUN PRESSFIT STR STEM12X110MM; KNEE STEMS Back to Search Results
Catalog Number 151312110
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Codes Bone Fracture(s) (1870); Pain (1994)
Date of Event 07/25/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that patient's primary knee was done.The patient began to have pain in tibia.Upon looking at xrays, the surgeon determined the tibial stem was through the lateral cortex.The patient underwent revision surgery.The entire tibial component and poly was removed, a total of five implants.Affected side: right knee.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: it was reported that patient's primary knee was done.Sales rep did not have information about the first revision that was done on (b)(6) 2019.The patient began to have pain in tibia.Upon looking at xrays, surgeon determined the tibial stem was through the lateral cortex.Xrays were attached.The patient scheduled revision surgery on (b)(6) 2025.The entire tibial component and poly was removed.A total of 5 implants.These implants and lots are listed and marked with a blue x on the first revision implant log from (b)(6) 2019.All unmarked implants were retained.The product was not returned to j&j medtech orthopaedics however photos were provided for review.See attachment (b)(4) scan (b)(6) 2025.The x-ray investigation revealed that the atun pressfit str stem12x110mm was in an improper position, appears to be mispositioned from the bone surface of the tibia, the tibial stem was through the lateral cortex.With the information provided is not possible to determine a potential cause at this moment.The mode of failure of the device is multi-factorial and consideration must be given to all other potential influences such as surgical process, patient variables (e.G.Activity level and use), anatomical considerations and patient changes over time.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.The overall complaint was confirmed as the observed condition of the atun pressfit str stem12x110mm would have contributed to the complained issue.Based on the investigation findings, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.H11 additional narrative: added: d10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATUN PRESSFIT STR STEM12X110MM
Common Device Name
KNEE STEMS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key22784619
Report Number1818910-2025-14046
Device Sequence Number18609264
Product Code JWH
UDI-Device Identifier10603295043553
UDI-Public(01)10603295043553(17)280801(10)J06G72
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
P830055
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number151312110
Device Lot NumberJ06G72
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/28/2025
Supplement Date Received by Manufacturer09/23/2025
Initial Report FDA Received Date08/12/2025
Supplement Report FDA Received Date09/24/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/17/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ATTUNE CRS RP INSRT SZ 4 14MM; ATUN PRESSFIT STR STEM12X110MM; ATUN TIB AUG UNIV SZ 3/4 5MM; ATUN TIB SLV M/L 29MM FULL POR; ATUNE REV RP TIB BASE SZ 4 CEM
Outcome Attributed to Adverse Event Required Intervention;
Patient Age60 YR
Patient SexFemale
-
-