| Catalog Number |
14-533041 |
| Medical Device Problem Codes |
Material Fragmentation (1261); Naturally Worn (2988)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Current information is insufficient to permit a valid conclusion about the cause of this event.A follow up report will be sent upon completion of the device evaluation.Reference reports 3012447612-2025-00278 through 3012447612-2025-00285.
|
| |
|
Event or Problem Description
|
|
It was reported that eight zyston inner shafts had become worn and are beginning to send shards of metal upon striking with mallet by the surgeon.There was no reported patient or surgical impact.This is report three of eight for this event.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: "device evaluation anticipated, but not yet begun" is erroneous and no longer applies.The device was evaluated.Corrections in h3.Additional information in h6: investigation type, findings, and conclusions.Device evaluation: the returned device matches the information in the complaint file and was examined.Visual inspection revealed no signs of damage.A functional inspection was performed with a zyston inserter and a zyston trial which revealed the rod was able to thread into each as expected.Additionally, a mallet was used to hit the proximal end of the rod, which revealed no signs of metal shards coming off.Root cause: a definitive root cause cannot be determined as the returned device was found to function as expected.Dhr review: the dhr was reviewed.There are no indications of manufacturing issues which would have contributed to this event and the device was likely conforming when it left highridge medical¿s control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
|
| |
|
Event or Problem Description
|
|
It was reported that eight zyston inner shafts had become worn and are beginning to send shards of metal upon striking with mallet by the surgeon.There was no reported patient or surgical impact.This is report three of eight for this event.
|
| |
|
Search Alerts/Recalls
|