|
The c111 analyzer serial number was (b)(6).The bilt3 reagent lot number used on the cobas 8000 system was 810295 with an expiration date of 31-jan-2026.The field service engineer (fse) cleaned the photometer, lens, and rotor.The fse checked various instrument parts and cleaned the water filter and purged the fluid system.Qc was performed.Patient samples were run and repeated for bilt3 and bild2 with acceptable results.The issue was not resolved.The fse returned to the customer site and decontaminated the fluid system and replaced the sample probe, and the photometer.A pipetting accuracy check was completed along with qc and patient sample monitoring.The investigation is ongoing.
|
|
Section d, device identification, was updated.Relevant fields of sections d and g were updated.The bilt3 reagent lot number was 815904 with an expiration date of 28-feb-2026.Over the course of multiple service visits, the following actions were performed by the field service engineer (fse): cleaned the 250 l syringe and replaced the teflon seal, cleaned the syringe linear guide, checked temperatures and fans, cleaned the photometer, lens, and rotor, checked the halogen lamp, sample syringe tip, and reagent, adjusted the sample probe, cleaned the water filter and gallon, and purged the fluid system, replaced the sample probe and the photometer, performed photometer initialization and air/water calibration, and checked pipetting accuracy.Afterward, the customer monitored patient samples.The customer stated the instrument was operating normally.The investigation determined that the service actions resolved the issue.As several service actions were performed, the specific cause of the event could not be determined.
|