| Catalog Number |
364966 |
| Medical Device Problem Code |
Loss of or Failure to Bond (1068)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/21/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported before using the bd vacutainer® urine transfer straw the cannula separated from the straw on an unspecified number of devices.No adverse impact was reported regarding the patient or user.
|
| |
|
Additional Manufacturer Narrative
|
|
The following fields were updated due to additional information: d9: device available for evaluation: yes.H3: device eval by manufacturer? yes.D9: returned to manufacturer on: 29-aug-2025.Investigation summary: bd received 200 samples and 6 photos for investigation.The samples and photos indicate that the holder has become separated from the straw and needle assembly.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.This complaint has been confirmed for the indicated failure mode separates.Regular inspections are performed on the adhesive bond to ensure that the bond will not break with a force of 5 lbs.No definitive root cause could be established for the reported defect.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for the identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported before using the bd vacutainer® urine transfer straw the cannula separated from the straw on an unspecified number of devices.No adverse impact was reported regarding the patient or user.
|
| |
|