| Model Number |
VTICMO_13.7 (- 15.00/4.0) |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Blurred Vision (2137)
|
| Date of Event |
07/03/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A facility representative reported that following an implantable collamer lens (icl) implantation procedure the patient experienced refractive surprise an unplanned post op femto-lasik to correct rx: -0,50/-0,50x9 occurred and the lens remains implanted.The cause of event was reported as unknown.This report is 1 of 2 total reports.Based on the information provided, the risk assessment for our product, and our experience, we have determined the cause or contribution to the reported issue is not indicative of a manufacturing related issue or product deficiency.
|
| |
|
Additional Manufacturer Narrative
|
|
H6 - type of investigation: 4110 - lens work order search: no similar complaint type events reported for units within the same lot.H11: manufacture narrative: refractive surprise each lens is subjected to a 100% assessment of the power (spherical and cylinder) and optical resolution during the manufacturing process in order to determine acceptability per the lens model and diopter.Claim #: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H3- supp#1 is not applicable.Claim #(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H3 - device evaluation: the lens was returned in liquid in a vial.Visual inspection found no visible damage to the lens.Dimensional and functional inspection found the lens to be within specifications.Claim # (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected data: d6b - in initial mdr is explant date: (b)(6) 2025.Claim#: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
B5- a facility representative reported that following an implantable collamer lens (icl) implantation procedure the patient experienced refractive surprise an unplanned post op femto-lasik to correct rx: -0,50/-0,50x9 occurred.The lens exchange for a same length lens and the problem resolved.The cause of event was reported as unknown.This report is 1 of 2 total reports.Based on the information provided, the risk assessment for our product, and our experience, we have determined the cause or contribution to the reported issue is not indicative of a manufacturing related issue or product deficiency.D10: injector model: msi-pf, lot # unk cartridge model: sfc-45, lot# unk foam tip plunger: ftp, lot# unk manufacture narrative: secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.Claim #(b)(4).
|
| |
|