• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC TORIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC TORIC INTRAOCULAR LENS Back to Search Results
Model Number VTICMO_13.7 (- 15.00/4.0)
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Blurred Vision (2137)
Date of Event 07/03/2024
Type of Reportable Event Serious Injury
Event or Problem Description
A facility representative reported that following an implantable collamer lens (icl) implantation procedure the patient experienced refractive surprise an unplanned post op femto-lasik to correct rx: -0,50/-0,50x9 occurred and the lens remains implanted.The cause of event was reported as unknown.This report is 1 of 2 total reports.Based on the information provided, the risk assessment for our product, and our experience, we have determined the cause or contribution to the reported issue is not indicative of a manufacturing related issue or product deficiency.
 
Additional Manufacturer Narrative
H6 - type of investigation: 4110 - lens work order search: no similar complaint type events reported for units within the same lot.H11: manufacture narrative: refractive surprise each lens is subjected to a 100% assessment of the power (spherical and cylinder) and optical resolution during the manufacturing process in order to determine acceptability per the lens model and diopter.Claim #: (b)(4).
 
Additional Manufacturer Narrative
H3- supp#1 is not applicable.Claim #(b)(4).
 
Additional Manufacturer Narrative
H3 - device evaluation: the lens was returned in liquid in a vial.Visual inspection found no visible damage to the lens.Dimensional and functional inspection found the lens to be within specifications.Claim # (b)(4).
 
Additional Manufacturer Narrative
Corrected data: d6b - in initial mdr is explant date: (b)(6) 2025.Claim#: (b)(4).
 
Additional Manufacturer Narrative
B5- a facility representative reported that following an implantable collamer lens (icl) implantation procedure the patient experienced refractive surprise an unplanned post op femto-lasik to correct rx: -0,50/-0,50x9 occurred.The lens exchange for a same length lens and the problem resolved.The cause of event was reported as unknown.This report is 1 of 2 total reports.Based on the information provided, the risk assessment for our product, and our experience, we have determined the cause or contribution to the reported issue is not indicative of a manufacturing related issue or product deficiency.D10: injector model: msi-pf, lot # unk cartridge model: sfc-45, lot# unk foam tip plunger: ftp, lot# unk manufacture narrative: secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.Claim #(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC TORIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
joselene muniz
1911 walker avenue
monrovia, CA 91016
MDR Report Key22786191
Report Number2023826-2025-01134
Device Sequence Number15517930
Product Code QCB
UDI-Device Identifier00840311333202
UDI-Public(01)00840311333202(11)240501(17)270430
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 08/12/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberVTICMO_13.7 (- 15.00/4.0)
Device Catalogue NumberVTICMO_13.7
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/18/2025
Supplement Date Received by Manufacturer08/18/2025
09/15/2025
10/24/2025
01/15/2026
Initial Report FDA Received Date08/12/2025
Supplement Report FDA Received Date09/12/2025
09/16/2025
11/18/2025
01/23/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/01/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age38 YR
Patient SexMale
-
-