SYNTHES GMBH LCK/NEUT GD 4.5 VA LCP CRVD COND AIM ARM; PLATE, FIXATION, BONE Back to Search Results
Catalog Number 03.231.007
Medical Device Problem Codes Device-Device Incompatibility (2919); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional narrative: d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.
 
Event or Problem Description
It was reported that during a distal femur procedure on an unknown date in 2025, the threaded locking towers were stripped along with bowling pin and other instruments.The surgeon wants all the hand drivers replaced, locking towers, trocars, neutral sleeves, cortical guides because they are extremely worn and do not snap into the aiming arm effectively.There was an unknown surgical delay.There was no patient consequence.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: added: d10.H6 component codes: most relevant component code is g07002 (appropriate term/code not available) to capture no findings available due to no product returned.H3, h6: product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LCK/NEUT GD 4.5 VA LCP CRVD COND AIM ARM
Common Device Name
PLATE, FIXATION, BONE
Manufacturer (Section D)
SYNTHES GMBH
luzernstrasse 19-21
zuchwil 4528
SZ  4528
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
elba bello
1302 wright lane east
west chester, PA 19380
9083863534
MDR Report Key22786651
Report Number8030965-2025-08291
Device Sequence Number15920084
Product Code HRS
UDI-Device Identifier10886982077349
UDI-Public(01)10886982077349
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K110354
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03.231.007
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/28/2025
Supplement Date Received by Manufacturer09/25/2025
Initial Report FDA Received Date08/12/2025
Supplement Report FDA Received Date09/25/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
HANDLE F/NOS. 324.203+324.215; LCK/NEUT GD 4.5 VA LCP CRVD COND AIM ARM; LCK/NEUT GD 4.5 VA LCP CRVD COND AIM ARM; LCK/NEUT GD 4.5 VA LCP CRVD COND AIM ARM; LCK/NEUT GD 4.5 VA LCP CRVD COND AIM ARM; LCK/NEUT GD 4.5 VA LCP CRVD COND AIM ARM; LCK/NEUT GD 4.5 VA LCP CRVD COND AIM ARM; LCK/NEUT GD 4.5 VA LCP CRVD COND AIM ARM; LCK/NEUT GD 4.5 VA LCP CRVD COND AIM ARM
Patient SexUnknown
-
-