| Model Number |
CF-Q180AL |
| Medical Device Problem Code |
Device Reprocessing Problem (1091)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/16/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation, olympus confirmed foreign material came out of the device, but the type of the material cannot be identified however, it is likely the following led to the malfunction: the foreign material was possibly not removed completely if the reprocessing steps were not conducted properly.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
No additional information provided by the customer.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was observed that during the device evaluation, the distal end (light guide rod lens) of the scope had foreign material.There was no patient involvement.
|
| |
|