Additional information provided in d.9., h.3., h.6., and h.11.No technical inquiries were received from the customer.One probe with tip protector in a tray with other items was received for the report.The sample was visually inspected and found to be nonconforming with orange/brown foreign material on the port face.White/clear foreign material was observed on the body needle around the port face.The sample was then functionally tested for actuation, aspiration and cut.The sample was found to be nonconforming for actuation, conforming for aspiration and nonconforming for cut.The probe was disassembled and the components inspected.Excessive wear observed on inner cutter when compared to the degree of wear based on continuous actuation of the probe visual standard photos.Gouge marks at bend area, cutting edge and several other areas on the inner cutter.A review of the device history record traceable to the lot number obtained from the device¿s radiofrequency identification (rfid) tag, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The investigation conducted a non-conformance review of the reported lot number.No deviations were identified, and all corresponding production release specifications defined in the device master record were met.Based on the evaluation of the information and materials received, the investigation was unable to identify the root cause or origin of the reported event.Additionally, no manufacturing-related deficiencies were found that potentially could have contributed to the complaint.The complaint evaluation does not confirm the probe had an aspiration failure; however, the observed actuation failure could have caused the cutter to stay in the port of the needle long enough to obstruct aspiration flow at the time the reported event was observed.The exact cause of the actuation failure cannot be determined from the evaluation performed.The complaint evaluation confirms that probe had a cut failure.The root cause for the poor cutting is the observed gouge marks.The probe was disassembled, and the components inspected.The wear observed on inner cutter when compared to the degree of wear based on continuous actuation of the probe visual standard photos is consistent with excessive use classification.Excess usage of the probe will wear and damage the inner cutter such that the cutter function becomes poor, bent, or does not function at all.Per the directions for use (dfu), the vitrectomy probe is verified for 20 minutes of use at maximum cut speed.As the root cause and its origin are inconclusive, further investigative or manufacturing actions are not warranted at this time as it appears that the observed cut failure was due to excessive use of the probe by the user.Per the directions for use (dfu), the vitrectomy probe is verified for 20 minutes of use at maximum cut speed.All probes are 100% visually inspected and tested for actuation, aspiration, and cut during manufacturing.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.Complaint data for all company products is reviewed monthly to monitor for adverse trends.During the last review, adverse trends were observed for the reported product and event combination.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
|