• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CURONIX LLC FREEDOM PERIPHERAL NERVE STIMULATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CURONIX LLC FREEDOM PERIPHERAL NERVE STIMULATOR Back to Search Results
Model Number FR8A-RCV-B0
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code Impaired Healing (2378)
Date of Event 07/16/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The surgical issue questionnaire was reviewed for potential causes of the reported issue.Based on this review, implanting the device after the expiration date and not prepping the skin with an antiseptic solution have been ruled out.However, the incision site was closed using staples which is non-compliant to the ifu.Per the freedom peripheral nerve stimulator system implantation of neurostimulator instructions for use, 05-20200 rev.7 "close the incisions using sterile skin closures and apply dressings.Closures are usually done in two steps, with absorbable sutures for the deeper layer and either nonabsorbable or absorbable sutures for the superficial layer." additionally, the patient has had trouble healing post-op in the past.A curonix representative conducted a review of sterilization and packaging records for the respective product lot; curonix has confirmed that the product was delivered sterile, validated sterilization parameters were used, and sterile barriers were verified to be intact following packaging.The stimulator is used to treat pain.The cause of the reported issue is attributed to two identified root causes: -non-compliance to the ifu as the incision site was closed using staples (user error-clinician).-patient is a poor candidate due to health, weight, age, or mental capacity as the patient has had trouble healing post-op in the past (user error-clinician).Rates reviewed at the most recent complaint trending meeting do not indicate a significant increase in surgical issues.Surgical issue rates remain acceptably low; thus, a capa is not required at this time.Surgical issue rates will continue to be tracked and trended.
 
Event or Problem Description
The patient reported impaired healing at the incision site.Antibiotics were prescribed and an explant procedure was performed on (b)(6) 2025.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREEDOM PERIPHERAL NERVE STIMULATOR
Common Device Name
PERIPHERAL NERVE STIMULATOR
Manufacturer (Section D)
CURONIX LLC
1310 park central boulevard s.
pompano beach FL 33064
Manufacturer (Section G)
CURONIX LLC
1310 park central boulevard s.
pompano beach FL 33064
Manufacturer Contact
alicia pagliaro
1310 park central boulevard s.
pompano beach, FL 33064
8009655134
MDR Report Key22789943
Report Number3010676138-2025-00164
Device Sequence Number18178947
Product Code GZF
UDI-Device Identifier00850051034314
UDI-Public(01)00850051034314(17)270101
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K171366
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 08/12/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberFR8A-RCV-B0
Device Lot NumberSWO250123
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/16/2025
Initial Report FDA Received Date08/12/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/23/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age62 YR
Patient SexFemale
-
-