|
This supplemental report is being submitted to provide the results of the final investigation.Updated fields: d9, h2, h3, h6, h11.The device was returned to olympus for inspection and the reported failure was not confirmed.In addition, the following reportable malfunctions were identified during the device evaluation: bent, dent and scratches on outer tube and image goes out of field.A root cause could not be identified.Based on the results of the investigation, it is likely the image goes out of field due to bent in outer tube.The most probable cause of bent, dent and scratches in outer tube is traced to component failure.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|