| Catalog Number |
5F060803CS |
| Medical Device Problem Codes |
Break (1069); Difficult to Remove (1528); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
07/18/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H11: as the lot number for the device was provided, a review of the device history records was performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.Section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2025, a patient underwent a stent placement procedure using lifestent 5f vascular stent.During the procedure, it was reported that the tip of deployment catheter broke off in patient upon stent deployment.Reportedly, it was unable to snare the broken part.The patient had to be taken to operating room and tip retrieved surgically.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2025, a patient underwent a stent placement procedure using lifestent 5f vascular stent.During the procedure, the tip of deployment catheter broke off in patient upon stent deployment.Reportedly, it was unable to snare the broken part.The patient had to be taken to operating room and tip retrieved surgically.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: manufacturing review: based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.However, the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Investigation summary: the returned sample is in used condition with activated deployment mechanism, and without stent that reportedly has been deployed inside patient.The inner catheter cardan tube is broken at the distal end, and the distal end of the cardan tube including tip, and both 1/4 rings is missing.The investigation leads to confirmed result for break and detachment of the inner catheter cardan tube.An indication for a manufacturing related cause could not be found.In this case only limited information was provided.Based on the information available, the investigation is closed with confirmed result for break and detachment of the inner catheter cardan tube.A definite root cause for the reported event could not be determined.Labeling review: in reviewing the relevant labeling for this product, the potential issue was found addressed.The instructions for use (ifu)states: 'do not constrict the delivery system during stent deployment.If excessive force is felt during stent deployment, do not force the stent system.Remove the stent system and replace with a new unit.' regarding pre dilation the ifu states: 'pre-dilatation of the lesion with a balloon dilatation catheter is recommended.' the instruction for use further state: 'gain ipsilateral or contralateral femoral access utilizing an appropriate introducer sheath ¿ 5f (1.67 mm) or larger introducer sheath.(.) insert a guidewire of appropriate length and 0.014 inch (0.36 mm) - 0.035 inch (0.89 mm) diameter (.)'.Regarding insertion and removal difficulty of the delivery system the ifu states: 'if resistance is met during stent system introduction, the stent system should be withdrawn and another stent system should be used.', and 'if resistance is met while retracting the delivery system over a guidewire, remove the delivery system and guidewire together.' holding and handling of the system throughout deployment was found sufficiently described.G2, g3, h6 (patient, component, method, result, conclusion).Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: manufacturing review: based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.However, the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Investigation summary: the returned sample is in used condition with activated deployment mechanism, and without stent that reportedly has been deployed inside patient.The inner catheter cardan tube is broken at the distal end, and the distal end of the cardan tube including tip, and both 1/4 rings is missing.Provided x-ray images demonstrate a short piece of catheter in the stented section inside the vessel.The investigation leads to confirmed result for break and detachment of the inner catheter cardan tube inside patient.An indication for a manufacturing related cause could not be found.In this case the vessel was tortuous and calcified, 6f introducer with.035" guidewire were used for access, the lesion was pre dilated, and a force increase was not felt during deployment and removal.The guidewire was running smoothly inside the system, the user did not experience difficulty during deployment, and the system was correctly held at the stability sheath during deployment action.The product was used off label in the iliac artery.Based on the information available, the investigation is closed with confirmed result for break and detachment of the inner catheter cardan tube.A definite root cause for the reported event could not be determined.The reported indication represents an off label use of the device.Labeling review: in reviewing the relevant labeling for this product, the potential issue was found addressed.The instructions for use (ifu) states: 'do not constrict the delivery system during stent deployment.If excessive force is felt during stent deployment, do not force the stent system.Remove the stent system and replace with a new unit.' regarding pre dilation the ifu states: 'pre-dilatation of the lesion with a balloon dilatation catheter is recommended.' the ifu further state: 'gain ipsilateral or contralateral femoral access utilizing an appropriate introducer sheath ¿ 5f (1.67 mm) or larger introducer sheath.(.) insert a guidewire of appropriate length and 0.014 inch (0.36 mm) - 0.035 inch (0.89 mm) diameter (.)'.Regarding insertion and removal difficulty of the delivery system the ifu states: 'if resistance is met during stent system introduction, the stent system should be withdrawn and another stent system should be used.', and 'if resistance is met while retracting the delivery system over a guidewire, remove the delivery system and guidewire together.' holding and handling of the system throughout deployment was found sufficiently described.The lifestent 5f vascular stent system is intended to improve luminal diameter in the treatment of symptomatic de novo or restenostic lesions up to 240 mm in length in the native superficial femoral artery (sfa) and popliteal artery.G3, h6 (method, conclusion).Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2025, a patient underwent a stent placement procedure using lifestent 5f vascular stent.During the procedure, the tip of deployment catheter broke off in patient upon stent deployment.Reportedly, it was unable to snare the broken part.The patient had to be taken to operating room and tip retrieved surgically.
|
| |
|