MAUDE Adverse Event Report: ANGIOMED GMBH & CO. MEDIZINTECHNIK KG LIFESTENT 5F VASCULAR STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIOMED GMBH & CO. MEDIZINTECHNIK KG LIFESTENT 5F VASCULAR STENT Back to Search Results
Catalog Number 5F060803CS
Medical Device Problem Codes Break (1069); Difficult to Remove (1528); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 07/18/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H11: as the lot number for the device was provided, a review of the device history records was performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.Section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
On (b)(6) 2025, a patient underwent a stent placement procedure using lifestent 5f vascular stent.During the procedure, it was reported that the tip of deployment catheter broke off in patient upon stent deployment.Reportedly, it was unable to snare the broken part.The patient had to be taken to operating room and tip retrieved surgically.
 
Event or Problem Description
On (b)(6) 2025, a patient underwent a stent placement procedure using lifestent 5f vascular stent.During the procedure, the tip of deployment catheter broke off in patient upon stent deployment.Reportedly, it was unable to snare the broken part.The patient had to be taken to operating room and tip retrieved surgically.
 
Additional Manufacturer Narrative
H11: manufacturing review: based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.However, the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Investigation summary: the returned sample is in used condition with activated deployment mechanism, and without stent that reportedly has been deployed inside patient.The inner catheter cardan tube is broken at the distal end, and the distal end of the cardan tube including tip, and both 1/4 rings is missing.The investigation leads to confirmed result for break and detachment of the inner catheter cardan tube.An indication for a manufacturing related cause could not be found.In this case only limited information was provided.Based on the information available, the investigation is closed with confirmed result for break and detachment of the inner catheter cardan tube.A definite root cause for the reported event could not be determined.Labeling review: in reviewing the relevant labeling for this product, the potential issue was found addressed.The instructions for use (ifu)states: 'do not constrict the delivery system during stent deployment.If excessive force is felt during stent deployment, do not force the stent system.Remove the stent system and replace with a new unit.' regarding pre dilation the ifu states: 'pre-dilatation of the lesion with a balloon dilatation catheter is recommended.' the instruction for use further state: 'gain ipsilateral or contralateral femoral access utilizing an appropriate introducer sheath ¿ 5f (1.67 mm) or larger introducer sheath.(.) insert a guidewire of appropriate length and 0.014 inch (0.36 mm) - 0.035 inch (0.89 mm) diameter (.)'.Regarding insertion and removal difficulty of the delivery system the ifu states: 'if resistance is met during stent system introduction, the stent system should be withdrawn and another stent system should be used.', and 'if resistance is met while retracting the delivery system over a guidewire, remove the delivery system and guidewire together.' holding and handling of the system throughout deployment was found sufficiently described.G2, g3, h6 (patient, component, method, result, conclusion).Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Additional Manufacturer Narrative
H11: manufacturing review: based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.However, the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Investigation summary: the returned sample is in used condition with activated deployment mechanism, and without stent that reportedly has been deployed inside patient.The inner catheter cardan tube is broken at the distal end, and the distal end of the cardan tube including tip, and both 1/4 rings is missing.Provided x-ray images demonstrate a short piece of catheter in the stented section inside the vessel.The investigation leads to confirmed result for break and detachment of the inner catheter cardan tube inside patient.An indication for a manufacturing related cause could not be found.In this case the vessel was tortuous and calcified, 6f introducer with.035" guidewire were used for access, the lesion was pre dilated, and a force increase was not felt during deployment and removal.The guidewire was running smoothly inside the system, the user did not experience difficulty during deployment, and the system was correctly held at the stability sheath during deployment action.The product was used off label in the iliac artery.Based on the information available, the investigation is closed with confirmed result for break and detachment of the inner catheter cardan tube.A definite root cause for the reported event could not be determined.The reported indication represents an off label use of the device.Labeling review: in reviewing the relevant labeling for this product, the potential issue was found addressed.The instructions for use (ifu) states: 'do not constrict the delivery system during stent deployment.If excessive force is felt during stent deployment, do not force the stent system.Remove the stent system and replace with a new unit.' regarding pre dilation the ifu states: 'pre-dilatation of the lesion with a balloon dilatation catheter is recommended.' the ifu further state: 'gain ipsilateral or contralateral femoral access utilizing an appropriate introducer sheath ¿ 5f (1.67 mm) or larger introducer sheath.(.) insert a guidewire of appropriate length and 0.014 inch (0.36 mm) - 0.035 inch (0.89 mm) diameter (.)'.Regarding insertion and removal difficulty of the delivery system the ifu states: 'if resistance is met during stent system introduction, the stent system should be withdrawn and another stent system should be used.', and 'if resistance is met while retracting the delivery system over a guidewire, remove the delivery system and guidewire together.' holding and handling of the system throughout deployment was found sufficiently described.The lifestent 5f vascular stent system is intended to improve luminal diameter in the treatment of symptomatic de novo or restenostic lesions up to 240 mm in length in the native superficial femoral artery (sfa) and popliteal artery.G3, h6 (method, conclusion).Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
On (b)(6) 2025, a patient underwent a stent placement procedure using lifestent 5f vascular stent.During the procedure, the tip of deployment catheter broke off in patient upon stent deployment.Reportedly, it was unable to snare the broken part.The patient had to be taken to operating room and tip retrieved surgically.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFESTENT 5F VASCULAR STENT
Common Device Name
VASCULAR STENT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM  76227
Manufacturer (Section G)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM   76227
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key22790799
Report Number9681442-2025-00245
Device Sequence Number8264406
Product Code NIP
UDI-Device Identifier00801741120176
UDI-Public(01)00801741120176(17)270517(10)ANJS1689
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
P070014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/30/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number5F060803CS
Device Lot NumberANJS1689
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/11/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/18/2025
Supplement Date Received by Manufacturer08/19/2025
09/30/2025
Initial Report FDA Received Date08/13/2025
Supplement Report FDA Received Date09/09/2025
10/27/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-