H3:product analysis:evaluation determined that motor separated from knuckle assembly.The likely cause of the failure is corrosion.It was also noted that exhaust hose found damaged, collet was jammed and corroded, depth test failed, nose flow test failed 3.2slpm and cosmetic condition of product is not okay.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|