• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOS SRL NUERA TIGHT 1.2; RADIO-FREQUENCY SKIN CONTURING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOS SRL NUERA TIGHT 1.2; RADIO-FREQUENCY SKIN CONTURING SYSTEM Back to Search Results
Model Number NUERA TIGHT 1.2
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Blister (4537)
Date of Event 07/23/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Lumenis investigated the reported event by contacting the user facility directly.Attempts by emails and phone calls have been made to obtain; patient treatment settings, patient information and patient photo.No incident form or patient photo has been received in lumenis.The device was installed on (b)(6) 2021 and is billable.No pm was performed since then based on device service history.Lumenis highly recommends an annual pm check in a timely manner to assure continued proper operation of the device.Initiative and payment for that lies on customer's responsibility.According to the technical support team, if the shielding is damaged or worn out in any way, it must be replaced by the user facility.The customer has not reported any device errors, and the system appears to be functioning properly.For the nuera 1.2 device, lumenis recommends using disposable grounding pads only.The technical support team has made several attempts to contact the customer to discuss the reported event and review technical data.However, the customer has been largely unresponsive.According to the ce, the customer continues to use the device without any issues.The ce advised transitioning from reusable grounding pads to disposable ones, but no response has been received to date.Device malfunction wasn't the suspected cause of the adverse event.The user manual includes the following instruction: "discard the single-use return plates after each patient use.Use only the single-use plates provided by lumenis." while the information received to date does not confirm that a serious injury nor malfunction had occurred, because of the lack of information lumenis is reporting the event in an 'abundance of caution'.Should additional significant information become available, the complaint record will be updated accordingly, and a follow-up medwatch report will be submitted.Lumenis is not the manufacturer of the nuera tight device which was used during this procedure, lumenis has forwarded the information of this event to the legal manufacturer and is closing this complaint.
 
Event or Problem Description
Lumenis received an adverse event report on a patient who sustained blister using reusable return plate following treatment by nuera device.According to the patient, the reusable grounding pad fel hot during the treatment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUERA TIGHT 1.2
Common Device Name
RADIO-FREQUENCY SKIN CONTURING SYSTEM
Manufacturer (Section D)
BIOS SRL
via guido rossa, 10/12
vimodrone, ita 20055
IT  20055
MDR Report Key22792969
Report Number3020611964-2025-00009
Device Sequence Number18303571
Product Code PBX
UDI-Device Identifier08052049500111
UDI-Public08052049500111
Combination Product (Y/N)N
Initial Reporter CountryIS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Importer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 08/13/2025
Report Date (Section F) 08/13/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Other
Device Model NumberNUERA TIGHT 1.2
Device Catalogue NumberGA-5200000US
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Event Location Outpatient Treatment Facility
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date08/13/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
Date Report Sent to FDA08/13/2025
-
-