|
The device was returned to stryker sustainability solutions for evaluation.Upon inspection of the received complaint device, the returned complaint device revealed three reprocessing dots.Inspection of the connector junction, revealed no separation.The connector appeared had no visual indication of damage.The tubing was discolored with no visual kinks, bends, or distress.The entirety of the cuff revealed no obvious signs of rips, tears, piling, fraying, or lacerations.Functional testing was performed using the bpc-ptc flow tester.The compressed air was set to 300.0mmhg for testing purposes and measured using calibrated manometer.The complaint device was connected to the device under testing (dut) tubing set and checked for hermeticity.The ball valve was then opened to allow for the complaint device to fill and stabilize.The cuff remained fully inflated with no visual or audible leaks seen/heard as indicated by a steady inflation rate of 300.0mmhg and a steady analog leak flow rate of 0.0mmhg.A review of the dhr supports that the device met all inspection and test criteria prior to release from stryker.The reported event could be attributed to: under pressurized pressure tourniquet cuff, wrong size tourniquet cuff used, and kink in tubing of ptc.The instructions for use (ifu) state: warnings: it is important that the tourniquet cuff be applied at the proper location with adequate pressure for the appropriate amount of time.Avoid needles, towel clips, leg holders and other equipment that can puncture or otherwise damage the cuff.Directions for use: before beginning the procedure, verify compatibility of all devices and accessories.Prior to surgery, select the proper sized tourniquet cuff by measuring the circumference of the patient's limb.This will avoid problems caused by a tourniquet cuff that is too small or too large.Secure the cuff fasteners to ensure that the cuff stays in place during the procedure.If the package is damaged or if it was opened and the device was not used, return the device and packaging to stryker sustainability solutions.Inspect the device for overall condition and physical integrity.Do not use the device if any damage is noted.Return the device and packaging to stryker sustainability solutions if it is not in acceptable condition for surgery.The reported event will continue to be monitored through post-market surveillance.
|