This medwatch submission is both an initial and supplemental filing.Investigation of the results can be seen below: investigation summary: no samples (including photos) were returned therefore the complaint could not be confirmed and the root cause is undetermined.This is the 1st complaint for the reported lot number.A review of the manufacturing records was performed, and no non-conformances were raised in association with this type of event for this lot.Complaints received for this device and reported condition will continue to be tracked and trended.If samples are received in the future the complaint will be reopened for further investigation.Based on the above, no additional investigation and no corrective/preventative action (capa) or situational analysis (sa) is required at this time.
|
|
Additional information was added to: b4, g6, h2, h3, h11.Correction to: h6 (component code, type of investigation, investigation findings, and investigation conclusions).Investigation summary: samples were received and an investigation was performed.This is the 1st complaint for the reported lot number.A review of the manufacturing records was performed and no non-conformances were raised in association with this type of event for this lot.Embecta was able to duplicate or confirm the indicated issue as bent non-patient end cannula and based on trend analysis no further action is required at this time.Complaints received for this device and reported condition will continue to be tracked and trended.As the samples were returned in open condition, the root cause for the bent non patient end cannula cannot be determined.Based on the above, no additional investigation and corrective/preventative action (capa) or situational analysis (sa) is required.
|