• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EMBECTA PENEL LIMITED - DUN LAOGHAIRE, IRELAND PEN NEEDLE AUTOSHIELD DUO 30GX5MM USA; NEEDLE, HYPODERMIC, SINGLE LUMEN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EMBECTA PENEL LIMITED - DUN LAOGHAIRE, IRELAND PEN NEEDLE AUTOSHIELD DUO 30GX5MM USA; NEEDLE, HYPODERMIC, SINGLE LUMEN Back to Search Results
Model Number 329515
Medical Device Problem Code Failure to Deliver (2338)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This medwatch submission is both an initial and supplemental filing.Investigation of the results can be seen below: investigation summary: no samples (including photos) were returned therefore the complaint could not be confirmed and the root cause is undetermined.This is the 1st complaint for the reported lot number.A review of the manufacturing records was performed, and no non-conformances were raised in association with this type of event for this lot.Complaints received for this device and reported condition will continue to be tracked and trended.If samples are received in the future the complaint will be reopened for further investigation.Based on the above, no additional investigation and no corrective/preventative action (capa) or situational analysis (sa) is required at this time.
 
Event or Problem Description
Consumer reported that needle was clogging during the injection lot # 4711012 catalog # 329515 date of event: unknown sample status: yes.(b)(6) 01august2025.Update for case # (b)(4).Lot # 4311012.
 
Additional Manufacturer Narrative
Additional information was added to: b4, g6, h2, h3, h11.Correction to: h6 (component code, type of investigation, investigation findings, and investigation conclusions).Investigation summary: samples were received and an investigation was performed.This is the 1st complaint for the reported lot number.A review of the manufacturing records was performed and no non-conformances were raised in association with this type of event for this lot.Embecta was able to duplicate or confirm the indicated issue as bent non-patient end cannula and based on trend analysis no further action is required at this time.Complaints received for this device and reported condition will continue to be tracked and trended.As the samples were returned in open condition, the root cause for the bent non patient end cannula cannot be determined.Based on the above, no additional investigation and corrective/preventative action (capa) or situational analysis (sa) is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PEN NEEDLE AUTOSHIELD DUO 30GX5MM USA
Common Device Name
NEEDLE, HYPODERMIC, SINGLE LUMEN
Manufacturer (Section D)
EMBECTA PENEL LIMITED - DUN LAOGHAIRE, IRELAND
Manufacturer Contact
anna garbowska-zakrzewski
300 kimball dr.
parsippany, NJ 07054
MDR Report Key22793316
Report Number3023359743-2025-00456
Device Sequence Number20960306
Product Code FMI
UDI-Device Identifier00382903295159
UDI-Public00382903295159
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K223286
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/11/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number329515
Device Lot Number4311012
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/31/2025
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date08/13/2025
Supplement Report FDA Received Date11/11/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/01/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Patient Sequence Number1
Patient SexUnknown
-
-