| Model Number |
MAJ-554 |
| Medical Device Problem Code |
Poor Quality Image (1408)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/14/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the camera head had poor image quality.The issue was found during a setting up the device before patient was inside the room.There were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
No additional information received from customer.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the final investigation.Updated fields: d8, h2, h3, h6, h11.The device was returned to olympus for inspection, and the reported failure was not confirmed.In addition, the following reportable malfunctions were identified during the device evaluation: image defect (noise) and damage to the headstock.A definitive root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: no problem detected.Olympus will continue to monitor field performance for this device.
|
| |
|