| Model Number |
3300TFX |
| Medical Device Problem Code |
Calcified (1077)
|
| Health Effect - Clinical Codes |
Cardiogenic Shock (2262); Heart Failure/Congestive Heart Failure (4446)
|
| Date of Event |
07/22/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a patient with a 29mm 3300tfx aortic valve underwent a valve-in-valve procedure after an implant duration of 4 years, 6 months due to calcification, degeneration, increased gradient, and severe stenosis.The patient presented with heart failure, dyspnea and fatigue.The tavr was performed with a 26mm 9755rsl transcatheter valve.The patient was in stable condition post procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: additional manufacturer narrative: the device was not returned for evaluation, as it remains implanted.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: additional manufacturer narrative: updated: a4, b4, b5, b7, g3, g6, h2, h6 (type of investigation).
|
| |
|
Event or Problem Description
|
|
It was reported and learned through investigation that a patient with a 29mm 3300tfx aortic valve underwent a valve-in-valve procedure after an implant duration of 4 years, 6 months due to calcification, degeneration, increased gradient, and severe stenosis.The patient presented with cardiogenic shock and acute heart failure.Nyha iv.Emergent salvage tavr procedure was performed with a 26mm 9755rsl valve.The patient was in stable condition post procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: additional manufacturer narrative: updated: b4, g3, g6, h2.H11: corrected data: h6 (health effect - clinical code, investigation conclusions), h11 (manufacturer narrative).Svd is a logical and expected consequence of the chemical, mechanical, and immunological processes associated with bhv implantation.The common endpoint of all these factors is calcification and degradation.Svd is an acquired condition, intrinsic to bioprosthetic valves, characterized by deterioration of the leaflets or supporting structures, leading to thickening, calcification, tearing, or disruption of the prosthetic valve materials.These changes result in valvular dysfunction manifesting as stenosis and/or regurgitation with a consequent drop in hemodynamic efficacy of the valve.Prior investigations and extensive literature review have shown that svd is predominantly related to patient factors and implant duration rather than to manufacturing issues.Events related to device design, manufacturing, or use error tend to manifest at an earlier implant duration.There is no evidence to suggest a manufacturing defect caused or contributed to the reported event.Based on the information available, the most likely root cause is patient factors, including hyperlipidemia, coronary artery disease.The device history record (dhr) review was completed and this device passed all manufacturing and sterilization inspections prior to release for distribution.There were no issues identified that would have impacted this event.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: additional manufacturer narrative: updated: b4, g3, g6, h2, h6 (type of investigation, investigation findings, investigation conclusions).Svd is a logical and expected consequence of the chemical, mechanical, and immunological processes associated with bhv implantation.The common endpoint of all these factors is calcification and degradation.Svd is an acquired condition, intrinsic to bioprosthetic valves, characterized by deterioration of the leaflets or supporting structures, leading to thickening, calcification, tearing, or disruption of the prosthetic valve materials.These changes result in valvular dysfunction manifesting as stenosis and/or regurgitation with a consequent drop in hemodynamic efficacy of the valve.Prior investigations and extensive literature review have shown that svd is predominantly related to patient factors and implant duration rather than to manufacturing issues.Events related to device design, manufacturing, or use error tend to manifest at an earlier implant duration.There is no evidence to suggest a manufacturing defect caused or contributed to the reported event.Based on the information available, the root cause remains indeterminable as no patient or procedural factors known to cause or contribute to svd were reported.The device history record (dhr) review was completed and this device passed all manufacturing and sterilization inspections prior to release for distribution.There were no issues identified that would have impacted this event.
|
| |
|