Catalog Number 3502700
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Codes Capsular Contracture (1761); Deformity/ Disfigurement (2360)
Date of Event 05/07/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
At the time of this report, mentor has received no information regarding explantation or an expected explantation date.It is unknown at this time if the device will be made available for return.As a result, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation is in progress.Once completed, a supplemental report will be submitted.Reason for device explant and/or reoperation: no information regarding explantation or an expected explantation date has been received.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a 52-year-old caucasian female patient underwent a breast surgery with 700cc mentor smooth round moderate plus profile saline breast prosthesis and experienced a deflation on the right side post-operatively.At the time of this report, mentor has received no information regarding explantation or an expected explantation date.
 
Additional Manufacturer Narrative
A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On september 24, 2025, mentor received additional information reporting that the patient was to undergo device explantation on (b)(6) 2025.Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.Reason for device explant and/or reoperation: deflation.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On (b)(6) 2025, mentor received additional information reporting that the patient also experienced capsular contracture (baker grade 4).The patient's replacement device was a 630cc mentor smooth round high profile saline breast prosthesis (catalog number 3503630) with lot number 9819461.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On november 10, 2025, the mentor failure analysis lab has received the device for evaluation.On november 15, 2025, the evaluation for the device was completed.Device evaluation summary: visual analysis of the returned device revealed that the breast implant had tear on the anterior view within a crease/fold, measuring approximately 1.6 cm.In addition, a crease/fold was noted from the posterior view extending to the anterior view.The evaluation determined that the possible cause of the rupture is consistent with normal wear.Deflation can occur at any time after implantation, but it is more likely to occur the longer the implant is implanted.Creating wrinkles or folds should be avoided during implantation or other procedures as it can result in a deflation in a later time.Breast implants may also simply wear out over time.Regarding the reported capsular contracture condition, there are not enough details to determine the factors that may have caused this condition.Capsular contracture in the patient¿s breast might be the result of the body¿s individual physiological response to the implantation of a foreign object in soft tissue.Further, capsular contracture is a known complication associated with these devices and is referenced in our current product insert data sheet as part of mentor's quality process, all devices are manufactured, inspected, and released according to approved specifications.Based on the results of this investigation, it was determined that the product met the manufacturing release criteria, and that normal wear may have caused the observed defect.No corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post market surveillance.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR SMOOTH ROUND MODERATE PLUS PROFILE
Common Device Name
PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
henry boden
3041 skyway circle north
irving, TX 75038
9497882078
MDR Report Key22797232
Report Number1645337-2025-08902
Device Sequence Number12902509
Product Code FWM
UDI-Device Identifier00081317001515
UDI-Public(01)00081317001515(17)200225(10)7298537
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P990075
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/25/2020
Device Catalogue Number3502700
Device Lot Number7298537
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/21/2025
Supplement Date Received by Manufacturer08/13/2025
09/24/2025
10/14/2025
11/10/2025
Initial Report FDA Received Date08/13/2025
Supplement Report FDA Received Date09/10/2025
10/10/2025
11/05/2025
11/18/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/26/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient Age52 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-