| Catalog Number |
0240031300 |
| Medical Device Problem Codes |
No Display/Image (1183); Output Problem (3005)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/15/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that during the surgery, the tower was functioning normally when suddenly the monitor turned off and did not display an image at any time afterward.After quickly checking possible solutions, the monitor still did not show an image, which led to an immediate replacement of the tower with another unit provided by the client.No adverse event occurred, although the procedure time was extended due to the equipment change.
|
| |
|
Event or Problem Description
|
|
It was reported that during the surgery, the tower was functioning normally when suddenly the monitor turned off and did not display an image at any time afterward.After quickly checking possible solutions, the monitor still did not show an image, which led to an immediate replacement of the tower with another unit provided by the client.No adverse event occurred, although the procedure time was extended due to the equipment change.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for investigation.This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: no live video output.Probable root cause: 1.Improper installation.2.Video cables are not sufficiently connected to ensure video transmission.3.Failure of edid negotiation.4.Failure of power supply.5.Failure or brownout of display voltage converter.6.User connects incompatible video source.7.Video source other than the one intended is selected for display.8.Damaged pins of hdmi/dvi connector.9.Open/short circuit failure between video connectors, tconn board and main board.10.Hard power switch on rear of display is in the off position.11.Failure/bug of scaler software.12.Incompatible cable used to connect video source to display.13.Cables have insufficient retention force.14.Cybersecurity event.15.Excessive usb current draw.16.Improper low voltage error setpoint.17.Open/short circuit of power cable caused by pinch force during assembly and shipment.18.Short circuit failure of tcon board electronics to tcon board cover (only on 0240031300 rev n and later).The reported failure mode will be monitored for future reoccurrence.
|
| |
|