| Medical Device Problem Code |
Patient-Device Incompatibility (2682)
|
| Health Effect - Clinical Codes |
Hypersensitivity/Allergic reaction (1907); Anaphylactoid / Anaphylaxis (2218)
|
| Date of Event |
09/30/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the complaint revealed that the patient wore the zio xt for 2 of the 3-day prescribed wear period.It is unknown if the patient has a history of reactions to medical adhesives or sensitive skin.The cause of the reported event cannot be determined.However, skin irritation is a known inherent risk of the device.The device manual warnings section read as follows: do not use the zio xt patch on patients with known allergic reaction to adhesives or hydrogels or with a family history of adhesive skin allergies.Patient may experience skin irritation.If skin irritation such as severe redness, itching or allergic symptoms develop, remove the zio xt patch from the patient¿s chest.This event is being reported per 21cfr803 as a serious injury.This report does not constitute an admission by rhythm that the product described in this report has any defects, or has malfunctioned.These terms are included in form fda 3500a and are fixed terms for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.
|
| |
|
Event or Problem Description
|
|
The patient presented to the urgent care with an anaphylactic allergic reaction and signs of a secondary infection allegedly caused by the zio xt.The patient reported that on day two, they experienced itching, redness, pain, and a rash.As a result, they removed the device and applied a medicated cream and benadryl at home.The patient¿s symptoms progressed over time, including trouble swallowing.While at the urgent care, the patient was administered a steroid shot, medications, and a cream.The patient is currently recovering.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental mdr is being submitted as a part of retrospective review from january 2021 till august 2025.A capa (2024-0036) was initiated on 29th september 2024 to implement corrective actions.Device performance and/or risk profile are not impacted.
|
| |
|
Search Alerts/Recalls
|