ERIKA DE REYNOSA, S.A. DE C.V. LIBERTY CYCLER SET, SINGLE PATIENT CONN; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
|
Back to Search Results |
|
| Catalog Number |
050-87215A |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/11/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A peritoneal dialysis (pd) patient reported the cycler prompted with m31 - air detected in cassette warnings during fill 5 of 5 of treatment.A fluid leak was discovered leaking from the patient line.Fluid was not discovered in the pump module area or or on the external of the cassette.The patient stated when the cycler warnings first occurred the patient touched the patient line and it felt a little bit wet and that was it.The cycler warnings occurred prior to the leak.The patient was advised to re-setup the cycler with all new supplies for the next treatment and to contact pd support if any issues occurred.Upon follow-up, the peritoneal dialysis registered nurse (pdrn) confirmed the patient's reported fluid leak event and stated a tubing leak was observed from a tubing line outside the cassette door.Fluid did not come in contact with the cycler and no fluid was noted within the pump module area, and the cassette remained dry.The pdrn stated the patient was trained on performing stat drains, as well as manual peritoneal dialysis therapy if needed and was able to drain manually and completed peritoneal dialysis treatment on the night of the reported event.The patient did not develop any symptoms, adverse events, injuries, or require medical intervention as a result of the reported event.The patient is continuing peritoneal dialysis therapy using the cycler with no further issues.The liberty cycler set used by the patient was discarded and was no longer available to return for physical evaluation by the manufacturer.
|
| |
|
Additional Manufacturer Narrative
|
|
Plant investigation: the plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.
|
| |
|
Additional Manufacturer Narrative
|
|
Plant investigation: as the device was not returned to the manufacturer, a physical evaluation could not be performed.A batch records review was conducted by the manufacturer for the reported lot.There were no non-conformances or abnormalities identified during the manufacturing process which could be associated with the reported event.The entire lot has been sold and distributed.In addition, a device history review was performed and confirmed that the results of the in-progress and final quality control (qc) testing met all requirements.The lot met all specifications for release.A product history review did not reveal a probable cause for the customer complaint.As a physical evaluation could not be performed, a definitive conclusion regarding the reported incident could not be reached and a cause could not be confirmed.
|
| |
|
Event or Problem Description
|
|
A peritoneal dialysis (pd) patient reported the cycler prompted with m31 - air detected in cassette warnings during fill 5 of 5 of treatment.A fluid leak was discovered leaking from the patient line.Fluid was not discovered in the pump module area or or on the external of the cassette.The patient stated when the cycler warnings first occurred the patient touched the patient line and it felt a little bit wet and that was it.The cycler warnings occurred prior to the leak.The patient was advised to re-setup the cycler with all new supplies for the next treatment and to contact pd support if any issues occurred.Upon follow-up, the peritoneal dialysis registered nurse (pdrn) confirmed the patient's reported fluid leak event and stated a tubing leak was observed from a tubing line outside the cassette door.Fluid did not come in contact with the cycler and no fluid was noted within the pump module area, and the cassette remained dry.The pdrn stated the patient was trained on performing stat drains, as well as manual peritoneal dialysis therapy if needed and was able to drain manually and completed peritoneal dialysis treatment on the night of the reported event.The patient did not develop any symptoms, adverse events, injuries, or require medical intervention as a result of the reported event.The patient is continuing peritoneal dialysis therapy using the cycler with no further issues.The liberty cycler set used by the patient was discarded and was no longer available to return for physical evaluation by the manufacturer.
|
| |
|