|
H11: as the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.Section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
H11: manufacturing review: the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.Investigation summary: the physical sample was not returned for evaluation.An angiographic image was provided showing two stents placed into each other.A stent deformation could not be identified.The image indicates a possible obstruction within the stent, but the cause of this obstruction could not be determined.Based on information available and evaluation of the provided footage, the investigation is closed with inconclusive result.A definite root cause for the reported issue could not be identified.Labeling review: relevant labeling for this product was reviewed.Potential complications and adverse events were found to be referenced.The instruction for use states that potential complications may include but are not limited to: new lesions in the access circuit requiring reintervention, thrombotic occlusion, restenosis of the target lesion requiring reintervention and others.Covered stent specific events that could be associated with clinical complications such as misplacement, migration, embolism, fracture, compression, kinking and insufficient covered stent expansion.Instructions for correct deployment of the covered stent were found to be described in the instruction for use.Regarding selection of the correct size of the covered stent the instruction for use states: "special care must be taken to ensure that an appropriately sized device is selected.In the case of a diameter difference between the inflow and the outflow end, utilize the following as the reference vessel depending on the type of access.For an av graft access, utilize the graft diameter and for an av fistula access, utilize the inflow vein diameter." regarding stent placement inside a previously placed stent the instruction for use states: "the device has not been tested for use in an overlapped condition with a bare metal stent or covered stent." e1, g2, g3, h6 (method).Section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|