| Model Number |
131F7 |
| Medical Device Problem Code |
Burst Container or Vessel (1074)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/09/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.One swan ganz catheter was received by our product evaluation laboratory for a full evaluation.As received, the balloon was found to be ruptured in the central area.The balloon edges did not appear to match up at the ruptured region.All through lumens were patent without any leakage or occlusion.Finally, no other visible damage was observed from catheter body.A supplemental report will be forthcoming with the engineering evaluation results.
|
| |
|
Event or Problem Description
|
|
As reported, during test of the balloon of this swan ganz catheter, there was a leakage.There was no allegation of patient injury.The device was available for evaluation.As per product evaluation, the balloon was found to be ruptured in the central area and the edges did not appear to match at the ruptured region.Patient demographics unable to be obtained.
|
| |
|
Additional Manufacturer Narrative
|
|
Updated section h6 (component code), h6 (impact code), h6 (type of investigation), h6 (investigation findings) and h6 (investigations conclusions).The manufacturing records were reviewed for the lot involved and there is no indication of a related nonconformance.All process parameters were met and inspections passed successfully.Based on the available information there is no evidence that supports or confirms the failure mode is associated to a manufacturing or design defect.As part of the manufacturing process the units go through balloon and bonding inspection process.
|
| |
|
Search Alerts/Recalls
|