• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES, PR SWAN-GANZ CATHETER; CATHETER, FLOW DIRECTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES, PR SWAN-GANZ CATHETER; CATHETER, FLOW DIRECTED Back to Search Results
Model Number 131F7
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/09/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.One swan ganz catheter was received by our product evaluation laboratory for a full evaluation.As received, the balloon was found to be ruptured in the central area.The balloon edges did not appear to match up at the ruptured region.All through lumens were patent without any leakage or occlusion.Finally, no other visible damage was observed from catheter body.A supplemental report will be forthcoming with the engineering evaluation results.
 
Event or Problem Description
As reported, during test of the balloon of this swan ganz catheter, there was a leakage.There was no allegation of patient injury.The device was available for evaluation.As per product evaluation, the balloon was found to be ruptured in the central area and the edges did not appear to match at the ruptured region.Patient demographics unable to be obtained.
 
Additional Manufacturer Narrative
Updated section h6 (component code), h6 (impact code), h6 (type of investigation), h6 (investigation findings) and h6 (investigations conclusions).The manufacturing records were reviewed for the lot involved and there is no indication of a related nonconformance.All process parameters were met and inspections passed successfully.Based on the available information there is no evidence that supports or confirms the failure mode is associated to a manufacturing or design defect.As part of the manufacturing process the units go through balloon and bonding inspection process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWAN-GANZ CATHETER
Common Device Name
CATHETER, FLOW DIRECTED
Manufacturer (Section D)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco
Manufacturer (Section G)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco
Manufacturer Contact
samantha eveleigh
1 edwards way
irvine, CA 92614
9492503939
MDR Report Key22802714
Report Number2015691-2025-06633
Device Sequence Number7836875
Product Code DYG
UDI-Device Identifier00690103000405
UDI-Public(01)00690103000405(17)250920(11)230921(10)65300129
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K233824
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/09/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date09/20/2025
Device Model Number131F7
Device Lot Number65300129
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/24/2025
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/12/2025
Supplement Date Received by Manufacturer09/08/2025
Initial Report FDA Received Date08/14/2025
Supplement Report FDA Received Date09/09/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/21/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-