MEDOS INTERNATIONAL SÃ RL HD LAPSCP,EP,5.5,0,300; LAPAROSCOPE, GENERAL & PLASTIC SURGERY
|
Back to Search Results |
|
| Catalog Number |
242632 |
| Medical Device Problem Codes |
Material Fragmentation (1261); Unintended Movement (3026)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/11/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11 additional narrative: the complaint device was received at the service center and evaluated.During the service evaluation the following defects were identified: other damages caused by customer, outer tube damaged, needle outer tube bent, outer tube damaged, distal tip distal tip damaged, illumination distal tip fiber damaged, particulate, optics particulate under distal lens particulate under proximal lens, optical system, optical components optical system loose prism loose cracked/broken negative, cosmetic issue loose eyepiece/coupler eyepiece.Visual: deformed/bent, broken (2+ pieces): device fractured, broken (2+ pieces): chipped/shaved/delaminated per service reports, this complaint was confirmed.Based on the investigation findings, the potential root cause is traced to user error.It has been determined that no corrective and/or preventative action is proposed and there is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
|
| |
|
Event or Problem Description
|
|
Itwas reported that during a laparoscopy surgical procedure; it was observed that the hd lapscp device was wobbly.There were no reports of delays to a surgical procedure.During in-house engineering evaluation, it was determined that the device was broken into two pieces - was chipped/shaved/delaminated.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11 additional narrative: the complaint device was received at the service center and evaluated.During the service evaluation the following defects were identified: other damages caused by customer, outer tube damaged, needle outer tube bent, outer tube damaged, distal tip distal tip damaged, illumination distal tip fiber damaged, particulate, optics particulate under distal lens particulate under proximal lens, optical system, optical components optical system loose prism loose cracked/broken negative, cosmetic issue loose eyepiece/coupler eyepiece.Visual : deformed/bent, broken (2+ pieces) : device fractured, broken (2+ pieces) : chipped/shaved/delaminated and usage : use error/misuse.Per service report, this complaint was confirmed.The faulty parts were replaced to resolve the issues.After repair, the device was found to be working according to the specifications.Based on the investigation findings, the potential root cause is traced to user error.It has been determined that no corrective and/or preventative action is proposed and there is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review manufacturing record evaluation is not required as the reported event is not associated with the manufacturing process and/or the potential cause of the defect cannot be associated to manufacturing.
|
| |
|
Search Alerts/Recalls
|
|
|