WILSON-COOK MEDICAL INC MEMORY EIGHT WIRE BASKET; BILIARY CATHETER AND ACCESSORIES Back to Search Results
Model Number G21910
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/04/2025
Type of Reportable Event Malfunction
Event or Problem Description
During an endoscopic retrograde cholangiopancreatography (ercp) in the bile duct, the physician used a cook memory eight wire basket.It was reported that the user opened the package and advanced the device through the endoscope working channel to desired position and discovered that a basket wire was broken.The user retracted the device and changed to a new one to complete the procedure.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Additional Manufacturer Narrative
E1: phone number: (b)(6).Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.A photo of the labeling was provided.The lot number provided in the photo matches this report.The label in the photo matches the product reported.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.Prior to distribution, all memory eight wire baskets are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: based on the quality engineering risk assessment no corrective action is warranted at this time.A review of the complaint history was conducted.Quality assurance will continue to monitor for complaint trends.
 
Additional Manufacturer Narrative
E1 - phone number: (b)(6).Investigation evaluation: the product said to be involved was returned in a clear plastic bag.Provided with the return was an open pouch from the lot number provided in the report.The label matches the product returned.A photo of the labeling was provided.The lot number provided in the photo matches this report.The label in the photo matches the product reported.Our laboratory evaluation of the product said to be involved confirmed the report.The device was returned with the basket fully retracted.The basket has one broken wire detached near the proximal end of the wire and remaining attached at the distal tip of the basket.The basket advanced and retracted as intended during handle manipulation with the broken wire remaining outside of the catheter.Under magnification of the broken wire evidence was found of embrittlement of the wire material.The fracture point of the basket wire was found to be close to the basket's proximal solder point, but there was no evidence of the wire being damaged by the buff process.A sticky brown substance was noted within the distal catheter and on the basket wires.No parts of the device appear to be missing.No other anomalies were detected.Photos were exchanged with production leadership and manufacturing engineering on 02sep2025.It was determined that the wire had been embrittled during the soldering process due to heating and reheating of the solder during the manufacturing process.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: our laboratory evaluation of the product said to be involved confirmed the report.The root cause of the broken basket wire was traced to the soldering process.Based on the visual examination it appears the basket wire was embrittled during the soldering process.This can occur during the heating and reheating of the solder and will not be visible to the operator.Production management and the department team leads were notified of this occurrence.The guidelines for soldering process has been revised to bring awareness to the risk of overheating during the soldering process and a retraining has been performed for all operators since the manufacture date of the affected lot to reduce occurrences of embrittlement.Prior to distribution, all memory eight wire baskets are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: based on the quality engineering risk assessment no corrective action is warranted at this time.A review of the complaint history was conducted.Quality assurance will continue to monitor for complaint trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEMORY EIGHT WIRE BASKET
Common Device Name
BILIARY CATHETER AND ACCESSORIES
Manufacturer (Section D)
WILSON-COOK MEDICAL INC
4900 bethania station rd
winston salem NC 27105
Manufacturer (Section G)
WILSON-COOK MEDICAL INC
4900 bethania station rd
winston salem NC 27105
Manufacturer Contact
sabrina o¿brien
4900 bethania station rd
winston salem, NC 27105
3367440157
MDR Report Key22803520
Report Number1037905-2025-00591
Device Sequence Number18407964
Product Code LQR
UDI-Device Identifier00827002219106
UDI-Public(01)00827002219106(17)280117(10)W4906057
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K171969
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/05/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG21910
Device Catalogue NumberMB5-3X6-8
Device Lot NumberW4906057
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/21/2025
Supplement Date Received by Manufacturer08/14/2025
Initial Report FDA Received Date08/14/2025
Supplement Report FDA Received Date09/08/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/17/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
OLYMPUS, TJF-260V.
Patient Age77 YR
Patient SexMale
Patient Weight55 KG
-
-