| Model Number |
CCA0T0 |
| Medical Device Problem Codes |
Failure to Unfold or Unwrap (1669); Failure to Deliver (2338)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/17/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
|
| |
|
Event or Problem Description
|
|
A non-healthcare professional reported that during an intraocular lens (iol) implant procedure, haptic was stuck and would not unfold.Additional information was received by stating, there was no patient contact.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information has been provided in d.9., h.3., h.6., and h.11.The product was returned for analysis and the reported complaint could not be observed.Iol (intraocular lens) was not returned.Only the device was returned.The device was returned loose in the carton.The lock-out assembly has been removed.Viscoelastic is observed in the device.The plunger has been fully advanced outside the nozzle.No damage observed.Iol was not returned.The product investigation could not identify the root cause for the reported complaint "haptic was stuck and would not unfold" as the lens was not returned.Only the device was returned for evaluation.Due to this, we are unable to confirm the lens and the plunger position for advancement and determine a root cause.Based on the results from the product history record, the products met release criteria.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
|
| |
|
Search Alerts/Recalls
|