• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR INC. GRAFTMASTER; CORONARY COVERED STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR INC. GRAFTMASTER; CORONARY COVERED STENT Back to Search Results
Model Number 1012581-19
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/22/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device was not returned for analysis.Production record and corrective and preventative actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a query of the complaint handling database for the reported lot revealed there is no indication of a lot specific issue.Based on the information provided, a definitive cause for the reported failure to advance could not be determined.Factors that could contribute to failure to advance include, but are not limited to, patient anatomical morphology, patient disease state, pre-dilatation strategy, device placement technique, interaction with previously placed stents or accessory devices.In this case, it is possible the device may have interacted with the patient anatomy during advancement, causing the reported failure to advance; however, based on the limited information provided by the account and without the device to examine, this cannot be confirmed.Based on the results of the complaint investigation, there is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.The additional device referenced in b5 is filed under a separate medwatch report number.
 
Event or Problem Description
It was reported the procedure was to treat a lesion in the left anterior descending artery.The 3.50x16mm rx graftmaster stent delivery system (sds) was advanced to cover a perforation however failed to cross the lesion.A 3.50x19mm and 2.80x26mm rx graftmaster also failed to cross the lesion.The procedure was completed using a 4.0x16mm graftmaster.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GRAFTMASTER
Common Device Name
CORONARY COVERED STENT
Manufacturer (Section D)
ABBOTT VASCULAR INC.
3200 lakeside drive
santa clara CA 95054 2807
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary UNK
EI   UNK
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key22803553
Report Number2024168-2025-08431
Device Sequence Number21070630
Product Code NIV
UDI-Device Identifier08717648176401
UDI-Public(01)08717648176401(17)250930(10)3102641
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
H000001
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/14/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/30/2025
Device Model Number1012581-19
Device Catalogue Number1012581-19
Device Lot Number3102641
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/24/2025
Initial Report FDA Received Date08/14/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/26/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age81 YR
Patient SexFemale
-
-