DEPUY IRELAND - 3015516266 EMSYS LNR IMP TIP 40; HIP INSTRUMENT - IMPACTOR Back to Search Results
Catalog Number 481440050
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/04/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that 40 impactor broke in half.Surgery time was not extended and all pieces were removed.
 
Additional Manufacturer Narrative
Product complaint #(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: d9.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary- it was reported that 40 impactor broke in half.Surgery time was not extended and all pieces were removed.The product was returned to j&j medtech orthopaedics for evaluation.Visual inspection of the returned device found the emsys lnr imp tip 40 is broken into two pieces.Both broken pieces were returned for evaluation.No other anomalies were noted.All fracture features observed are consistent with overload fracture.The fracture initiated and progressed due to overload, whereby a load was applied exceeding the ultimate strength of the material.The evidence suggests that the mating instrument was fully threaded/engaged, or minimally loose.A functional test was not performed since it was not applicable to the complaint condition.A dimensional inspection was not performed since it was not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the emsys lnr imp tip 40 would have contributed to the complained issue.Based on the investigation findings, the potential cause is traced to unintended use error and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot- the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.H11 additional narrative: corrected: h3.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMSYS LNR IMP TIP 40
Common Device Name
HIP INSTRUMENT - IMPACTOR
Manufacturer (Section D)
DEPUY IRELAND - 3015516266
loughbeg ringaskiddy co.
cork P43 E D82
EI  P43 ED82
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key22803848
Report Number1818910-2025-14174
Device Sequence Number12973379
Product Code HWA
UDI-Device Identifier10603295510208
UDI-Public(01)10603295510208(10)AZ499599
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/14/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number481440050
Device Lot NumberAZ499599
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/04/2025
Supplement Date Received by Manufacturer08/21/2025
10/01/2025
Initial Report FDA Received Date08/14/2025
Supplement Report FDA Received Date08/22/2025
10/02/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/12/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age72 YR
Patient SexMale
-
-