| Catalog Number |
MWM054 |
| Medical Device Problem Codes |
Break (1069); Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/17/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The glenosphere impactor tip was damaged during impaction.
|
| |
|
Additional Manufacturer Narrative
|
|
Once the investigation has been completed any additional information will be reported in a supplemental report.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected data: g3 (reportability awareness date).The reported event could not be confirmed, since the device was not returned for evaluation and no other evidences were provided.The device inspection was not possible as the product was not returned for investigation.The device history record could not be reviewed because the affected device was not returned, and the lot number was not communicated.Indications of material, manufacturing, or design related problems were unable to be identified as the lot number was not communicated.A review of the labeling did not indicate any abnormalities.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.A design improvement is in progress to prevent the recurrence of the alleged failure.If device is returned or any further information is provided, the complaint report will be updated.
|
| |
|
Event or Problem Description
|
|
The glenosphere impactor tip was damaged during impaction.
|
| |
|