| Model Number |
2C6255 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that an unspecified quantity of clearlink system continu-flo solution sets leaked from "port 4" (clearlink above the stopcock).The issue occurred during patient use when the bag was spiked with lactate ringers.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was added to h4, h6, and h11: h11: the actual device was not available; however, seventy (70) companion samples were received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.A pressure test and clear passage under water were performed, and no leaks were observed.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|