MDT PUERTO RICO OPERATIONS CO, MED REL INFINITY¿ OCCIPITOCERVICAL UPPER THORACIC SYSTEM; APPLIANCE, FIXATION, SPINAL INTERLAMINAL Back to Search Results
Model Number G3600215
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/25/2025
Type of Reportable Event Serious Injury
Event or Problem Description
Information was received from healthcare provider (hcp) via a manufacturer representative regarding a patient having c3-th1 posterior fixation therapy for c5/6 fracture.It was reported that the set screw detached two weeks after surgery without causing symptoms.Subsequent monitoring revealed two additional dislodgements, leading to a planned repeat surgery (on (b)(6) 2025).The physician attributed the issue to a product defect, not misuse.There was no patient symptom reported.There were no further complications reported regarding the event.Re-operation for replacing the rod and set screw was performed.Even the part where the set screw was not loosened could be removed easily (without applying more force than usual).Reoperation was performed due to set screw dislodgement after the initial surgery on (b)(6) 2025.
 
Additional Manufacturer Narrative
E1: first name and last name of initial reporter is unknown.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
H3: product analysis of product:g3600215, lot:h6040083 visual and functional testing identified that the threads of the set screw are damaged.This is consistent with misalignment.B2: outcome attributed to adverse event is additional surgery due to instrument malfunction.B5: event problem description is updated medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Additional information was received from the manufacturer representative that there were totally 8 products were involved in this event.
 
Additional Manufacturer Narrative
Radiographic image review states, that the lateral xray at c3 - t1 shows posterior fusion screws in c34, c7t1 and 2 panel x-ray shows left antero-posterior right lateral, the sent screw is out of the tulip on one side at c3.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFINITY¿ OCCIPITOCERVICAL UPPER THORACIC SYSTEM
Common Device Name
APPLIANCE, FIXATION, SPINAL INTERLAMINAL
Manufacturer (Section D)
MDT PUERTO RICO OPERATIONS CO, MED REL
road 909, km. 0.4., barrio mar
humacao PR 00792
Manufacturer (Section G)
MDT PUERTO RICO OPERATIONS CO, MED REL
road 909, km. 0.4., barrio mar
humacao PR 00792
Manufacturer Contact
justin ellis
1800 pyramid place
memphis, TN 38132
7635265677
MDR Report Key22809108
Report Number2647346-2025-00057
Device Sequence Number20783073
Product Code KWP
UDI-Device Identifier00763000474010
UDI-Public(01)00763000474010(17)330318(10)H6040083
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K163375
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/28/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG3600215
Device Catalogue NumberG3600215
Device Lot NumberH6040083
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/25/2025
Supplement Date Received by Manufacturer09/24/2025
10/21/2025
Initial Report FDA Received Date08/14/2025
Supplement Report FDA Received Date10/15/2025
10/28/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/18/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-