| Model Number |
8884720205E |
| Medical Device Problem Code |
Unintended Deflation (4061)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/24/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported that two devices have holes on the bulb.On (b)(6) 2025 the initial reporter confirmed the issue was with the balloon and occurred du ring patient use.No harm noted.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation is currently underway.Upon completion, the results will be forwarded.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record (dhr) could not be reviewed as no lot information was provided within the complaint.Prior to a lot¿s release, the lot must be deemed acceptable by passing inspections that are based on a valid sampling plan.During production, inspectors routinely examine a statistical sample both physically and visually.There were no samples returned for evaluation; therefore, the affected device could not be evaluated to confirm the reported failure mode.As such, a root cause could not be determined at this time.However, a supplier corrective action request has been sent to the supplier to further investigate and address the reported condition.We will continue to monitor related reports to determine if additional actions are necessary.
|
| |
|