• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDINAL HEALTH, INC. COVIDIEN; GASTROINTESTINAL TUBES WITH ENTERAL SPECIFIC CONNECTORS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDINAL HEALTH, INC. COVIDIEN; GASTROINTESTINAL TUBES WITH ENTERAL SPECIFIC CONNECTORS Back to Search Results
Model Number 8884720221E
Medical Device Problem Code Unintended Deflation (4061)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/24/2025
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported that bulbs are deflating because of holes.
 
Additional Manufacturer Narrative
An investigation is currently underway.Upon completion, the results will be forwarded.
 
Additional Manufacturer Narrative
The device history record was reviewed and indicated that the product was released accomplishing all quality standards.Prior to a lot¿s release, the lot must be deemed acceptable by passing inspections that are based on a valid sampling plan.During production, inspectors routinely examine a statistical sample both physically and visually.There were no samples returned for evaluation; therefore, the affected device could not be evaluated to confirm the reported failure mode.As such, a root cause could not be determined at this time.However, a supplier corrective action request has been sent to the supplier to further investigate and address the reported condition.We will continue to monitor related reports to determine if additional actions are necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COVIDIEN
Common Device Name
GASTROINTESTINAL TUBES WITH ENTERAL SPECIFIC CONNECTORS
Manufacturer (Section D)
CARDINAL HEALTH, INC.
3651 birchwood drive
waukegan IL 60085
Manufacturer (Section G)
CARDINAL HEALTH
calle 9 sur no. 125 cuidad ind
tijuana 22500
MX   22500
Manufacturer Contact
jill saraiva
3651 birchwood drive
waukegan, IL 60085
5086183640
MDR Report Key22812562
Report Number1423537-2025-00327
Device Sequence Number21008387
Product Code PIF
UDI-Device Identifier10884521661721
UDI-Public(01)10884521661721(17)270301(10)2333917264
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K180622
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/29/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number8884720221E
Device Catalogue Number8884720221E
Device Lot Number2333917264
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/25/2025
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date08/14/2025
Supplement Report FDA Received Date08/29/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/09/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-