PTS DIAGNOSTICS CARDIOCHEK PLUS ANALYZER; CLINICAL CHEMISTRY ANALYZER Back to Search Results
Model Number 1.12
Medical Device Problem Code Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/15/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
During the investigation, it is determined that the terminal displacement is most likely caused by the customer providing undue force onto the battery terminals.In addition, the potential exists for the battery contact to lose traction with the compartment causing less force to dislodge however improper removal of batteries will still force the dislodge.Several key factors point to this decision: the battery terminals would have to have been properly seated at the time of manufacturing because the analyzer would not power on and would consequently be unable to be calibrated and released.Upon close inspection of the battery terminal area, it was observed that the terminals were fully seated initially due to visible drag marks created by the metal terminal scraping the plastic casing when excessive force was applied.These markings are present from the point of terminal seating down to where the terminals lie currently.Pts is confident that this damage occurred post-distribution, as there are no manufacturing steps that would cause this type of damage.
 
Event or Problem Description
There were no allegations of patient or user harm.The customer reported that upon battery installation, their analyzer's battery compartment "got really hot and brand-new batteries exploded after removing them from the analyzer.No one was injured.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOCHEK PLUS ANALYZER
Common Device Name
CLINICAL CHEMISTRY ANALYZER
Manufacturer (Section D)
PTS DIAGNOSTICS
4600 anson blvd
whitestown IN 46075
Manufacturer (Section G)
PTS DIAGNOSTICS
4600 anson blvd
whitestown IN 46075
Manufacturer Contact
randall cundiff
4600 anson blvd
whitestown, IN 46075
MDR Report Key22812871
Report Number1836135-2025-00012
Device Sequence Number18332021
Product Code CGA
UDI-Device Identifier00381932701013
UDI-Public(01)00381932701013(8012)1.12
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K140068
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/14/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1.12
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/24/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/24/2025
Initial Report FDA Received Date08/14/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-