• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. THREE PEG PATELLA 29MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. THREE PEG PATELLA 29MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 200-02-29
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Codes Pain (1994); Joint Laxity (4526); Swelling/ Edema (4577)
Date of Event 03/30/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Concomitant products: - three peg patella 29mm (cat# 200-02-29 / serial# (b)(6)).The revision reported was likely due to the reported instability, or due to the inclusion of the polyethylene tibial insert in the packaging recall.Prosthesis wear of the patella appears to be an incidental finding during revision and does not appear to be the cause for the revision.Additional reasons for the reported revision may be tibial insert prosthesis wear and instability, however, the reported instability and tibial insert prosthesis wear could not be confirmed from the provided information.Additionally, a contributing factor to the reported wear may have been the result of the tibial insert being packaged in a non-conforming vacuum bag for more than five years.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and radiographs and relevant clinical information were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Event or Problem Description
As reported, the female patient had an original exactech left tka.The patient presented back to surgeon's office with complaints of pain, swelling, instability, and dissatisfaction with their left tka.The patient has a recalled implant.Upon examination, the patient was scheduled to have a revision tka possible poly swap vs full revision.The patient had revision surgery on (b)(6) 2023.The patient underwent a left tka tibial poly swap and patella implant swap and revised to a truliant crc insert size 1.5, 13mm and optetrak 3 peg patella cemented 29m.The patient left the operating room (or) stable.There was no breakage of a device or surgical delay/prolongation.Patient was last known to be in stable condition following the event.Images received.The devices are not available for evaluation due the implant was required to be sent to the lab and legal at the hospital.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THREE PEG PATELLA 29MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH INC.
2320 nw 66th ct
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th ct
gainesville, FL 32653
MDR Report Key22814268
Report Number1038671-2025-02762
Device Sequence Number9014345
Product Code JWH
UDI-Device Identifier10885862039583
UDI-Public10885862039583
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K932690
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/15/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/04/2021
Device Catalogue Number200-02-29
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/31/2023
Initial Report FDA Received Date08/15/2025
Date Device Manufactured10/06/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11.
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexFemale
-
-