ETHICON INC. PRLNE BLU 36IN 5-0 D/A BB; SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE Back to Search Results
Catalog Number 8580H
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/27/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint #(b)(4).H6 component code: g07002: device evaluation pending.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.The device upon which this medwatch is based has been received, however, the product evaluation is not yet complete.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Additional information was requested, and the following was obtained via: please provide additional details about the alleged deficiency.The needles got loosened when passing through the prostheses.How long was the delay? please specify in minutes.20 minutes.Were there any unexpected outcomes or complications as a result of the prolonged surgery time? i don't believe, not at the time of surgery, postoperatively.Was there any change in the patient¿s post-operative care due to the prolonged procedure? unknown.Were there patient consequences? if yes, please explain.Not at the time of surgery, postoperatively.How many devices demonstrated the alleged deficiency? 4 threads of the same lot number.It was reported that the event date is june 27, 2025, and the alert date is july 21, 2025.Could you please provide a justification for this variance in the timing of the report related to this files? we wrote the statement on the same day of the event i don't know why there was an alert until 1 months later.Please provide the date of event.27 june.Both incidents took place during the same procedure (1 patient concerned).
 
Event or Problem Description
It was reported that a patient underwent a cardiovascular procedure on (b)(6) 2025 and suture was used.During the procedure, at the time of suture between the valve prosthesis and cardiac prosthesis.The suture between the cardiorotic prosthesis and the aorta the 4 wires used have loosened risk of tearing during suture delay.No adverse patient consequences were reported.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).Date sent to the fda: 9/18/2025.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H6 component code: g07002 no device problem found.Additional information: d9, h3, h6.H3 investigational summary: the product was returned to ethicon for evaluation.Ten representative unopened samples were received for analysis.Product code 8580h.The complaint sample was not received for evaluation.Upon initial inspection of the samples, no external damages were observed on the external packets.In order to evaluate the conditions of the returned samples, the packets were opened, and no defects were detected.The swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along the strand to detect any issue related no defects were observed during evaluation.A functional test was performed using instron equipment, and the pull force results were above the minimum requirements.The event described could not be confirmed as the device performed without any defect noted.It should be noted that as part of our quality process, each batch is randomly inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.The event described could not be confirmed as no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Additional information was requested, the following was obtained: there were no post-operative consequences for the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRLNE BLU 36IN 5-0 D/A BB
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-SAN LORENZO PR
road 183, km 8.3
san lorenzo PR 00754
Manufacturer Contact
orla o'mahony
1000 route 202
raritan, NJ 08869
3035526892
MDR Report Key22814497
Report Number2210968-2025-09373
Device Sequence Number15706722
Product Code GAW
UDI-Device Identifier10705031017337
UDI-Public(01)10705031017337(17)47238(10)1011RL
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K133356
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/15/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number8580H
Device Lot Number1011RL
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/14/2025
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/21/2025
Supplement Date Received by Manufacturer08/21/2025
Initial Report FDA Received Date08/15/2025
Supplement Report FDA Received Date09/18/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/13/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-