• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TRIVIDIA HEALTH INC PEN NEEDLES FG; HYPODERMIC SINGLE LUMEN NEEDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TRIVIDIA HEALTH INC PEN NEEDLES FG; HYPODERMIC SINGLE LUMEN NEEDLE Back to Search Results
Model Number PNDL, 5BV TVH TP 32G 4MM100CT 30/CS
Medical Device Problem Code No Flow (2991)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/22/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal report reference number: (b)(4).Pen needles were not returned for evaluation.Note: manufacturer contacted customer in several follow-up calls to ensure the replacement products resolved the initial concern - unable to establish contact with customer at this time.
 
Event or Problem Description
Consumer reported complaint for the trueplus pen needles, stating that the 32g pen needles were not aspirating.The package was not opened or damaged when received by the customer.The customer has been using the product for 1 week.The customer is using compatible product, and the pen needle is properly aligned.At the time of the call the customer felt well and did not report any symptoms.Customer did not claim to be injured while using the pen needles and no medical intervention related to the use of the product was reported.
 
Additional Manufacturer Narrative
Sections with additional information as of 27-oct-2025: d9: device available for evaluation.H3: device evaluated by manufacturer.H6: updated fda¿s type, findings and conclusions codes.H10: pen needles were returned.Unable to ship returned product to the manufacturer due to biohazard.Return product scrapped.Complaint was forwarded to supplier quality and internal evaluation was performed by the manufacturer using pen needles from the same lot.No abnormalities observed with retention samples.Most likely underlying root cause: mlc-009: use error caused or contributed to event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PEN NEEDLES FG
Common Device Name
HYPODERMIC SINGLE LUMEN NEEDLE
Manufacturer (Section D)
TRIVIDIA HEALTH INC
2400 nw 55th court
fort lauderdale FL 33309
Manufacturer Contact
juliana arias
2400 nw 55th court
fort lauderdale, FL 33309
MDR Report Key22814632
Report Number1000113657-2025-00275
Device Sequence Number20262059
Product Code FMI
UDI-Device Identifier00021292010805
UDI-Public(01)00021292010805(17)291130(10)4L616
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K180720
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial,Followup
Report Date (Section B) 10/27/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberPNDL, 5BV TVH TP 32G 4MM100CT 30/CS
Device Lot Number4L616
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Distributor Facility Aware Date07/22/2025
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 07/22/2025
Supplement Date Received by Manufacturer10/23/2025
Initial Report FDA Received Date08/15/2025
Supplement Report FDA Received Date10/27/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-