| Model Number |
FLIPCUTTER II, SHORT 8.5MM |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/30/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Event or Problem Description
|
|
It was reported that during an anterior cruciate ligament surgery the tip of the flip cutter broke off during drilling.All broken parts were retrieved from the patient.Per complaint reporter there was no harm for patient, operator or third party.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6.One unpackaged ar-1204as-85, flipcutter¿ ii, batch 4208151303, was received for evaluation.- upon visual inspection, it was noted that the tip of the device was warped.- functional testing could not be performed due to the damage to the tip.- the most likely cause for the reported failure can be attributed to is misuse due to misaligned insertion and prying/leveraging the device during use.- the complaint allegation is confirmed.- refer to the investigation photos.
|
| |
|