Model Number DCB00
Medical Device Problem Codes Crack (1135); Mechanical Jam (2983)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/21/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Section a2, a4 and a5: per regulation eu 2016/679 (general data protection regulation), patient identifiers were not collected or recorded and therefore are not available.Section d6a - implant date: not applicable, as there is no indication that the lens was implanted.Section d6b - explant date: not applicable, as there is no indication that the lens was implanted.Section e1 - telephone number: (b)(6).Section h3 - the device was not returned for evaluation; therefore, a failure analysis of the complaint device cannot be completed.A review of records including device history and complaint trending will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported with the preloaded monofocal intraocular lens (iol), there was a defect in the injector tip.This caused the lens to protrude from the side of the device and rendering them unusable.During follow-up, it was confirmed that the material came into contact with the patient, but the customer was unable to remove the lens from the cartridge.No further information provided.
 
Additional Manufacturer Narrative
Additional information: section d9 device available for evaluation? yes.Section d9 date returned to manufacturer: october 9, 2025.Section h3 device evaluated by manufacturer? yes.Device evaluation: visual inspection under magnification was performed on the suspect product.The plunger rod was found advanced to a lens that was stuck in the cartridge.Viscoelastic residue was not observed to be evenly distributed throughout the cartridge indicating that an inadequate amount may have been used, which could contribute to the complaint issue reported.Stress marks were observed on the cartridge tip; the cartridge tip and cartridge were also observed to be damaged and bulging where the lens was stuck.No issues were observed with the lens module, device assembly, or plunger rod advancement.The plunger rod tip was observed to be slightly bent.The lens could not be removed from the cartridge for evaluation; however, it was observed to be damaged.No further evaluation was performed.The complaint issue "stuck in cartridge" and "cartridge tip cracked/damaged" were identified during product evaluation; however, based on the complaint investigation results the complaint issue could not be confirmed to be related to the manufacturing or design process.Therefore, there is no indication of a product deficiency or product malfunction.Conclusion: as a result of the investigation, there is no indication of a product quality deficiency.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SIMPLICITY
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key22814932
Report Number3012236936-2025-000207
Device Sequence Number5738797
Product Code HQL
UDI-Device Identifier05050474636170
UDI-Public(01)05050474636170(17)271217
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberDCB00
Device Catalogue NumberDCB0000245
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/21/2025
Supplement Date Received by Manufacturer11/12/2025
Initial Report FDA Received Date08/15/2025
Supplement Report FDA Received Date12/02/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/17/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-