Catalog Number 204-04-32
Medical Device Problem Codes Unstable (1667); Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Unspecified Infection (1930); Joint Laxity (4526); Swelling/ Edema (4577)
Date of Event 03/16/2022
Type of Reportable Event Serious Injury
Event or Problem Description
As reported via legal documentation, a patient had right knee replacement surgery due to osteoarthritis.They underwent right knee revision surgery approximately 8 years 8 months post primary procedure.Operative report operative findings: gross purulence and infection, loose femoral component, loose tibial component, well-fixed patellar component, adequate soft tissue coverage, aori class femur 2b, aori class tibia 2b.Postoperative diagnosis: pyogneic arthritis of right knee joint, due to unspecified organism.The surgeon noted abundant purulence.The femoral component was noted to be loose and completely debonded from the bone, where abundant purulence was noted in the femur around the implant, which was removed.The polyethylene was removed and the tibial component was also noted to be loose with severe bone loss and purulence in the bone.The patellar bone had osteolysis as well over the lateral facet.She tolerated the procedure well without complication.There is no other patient demographic or medical history available.There is no device return.There are no photos or other images of the device provided.No additional information is available.
 
Additional Manufacturer Narrative
Concomitants: optetrak tibial insert & screw constrained condylar (208-23-11, (b)(6), optetrak 3 peg patella (200-02-35, (b)(6), optetrak fluted stem extension femoral and tibial (204-32-01, (b)(6), optetrak tibial tray (204-04-32, (b)(6), optetrak femoral component non-modular constrained (210-01-03, (b)(6).The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear, instability, failure of the cement used to secure the femoral component to the femoral bone, which led to loosening at the cement-implant interface, tibial loosening, and infection or due to inclusion of the polyethylene in the packaging recall.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images and radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRAPEZOID TIBIAL TRAY SZ 3F/2T
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH INC.
2320 nw 66th ct
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th ct
gainesville, FL 32653
MDR Report Key22816073
Report Number1038671-2025-02766
Device Sequence Number5713103
Product Code JWH
UDI-Device Identifier10885862047656
UDI-Public10885862047656
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K933610
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/15/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/27/2014
Device Catalogue Number204-04-32
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/03/2023
Initial Report FDA Received Date08/15/2025
Date Device Manufactured12/29/2009
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11.
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age68 YR
Patient SexFemale
-
-